4.3
Manual Cleaning and disinfection   
Prepare following items before manual cleaning and disinfection:   
a) Detergent: Cidezyme (Johnson & Johnson, #2258) or another cleaning agent with 
approved material compatibility for this medical device 
b) Disinfectant: Cidex OPA (Johnson & Johnson, # 20391) or another disinfectant with 
approved material compatibility for this medical device 
c) Two tanks, one for cleaning and one for disinfection ‐ optional: 
1 additional tank for rinsing with deionized/tap water (sufficient size for immersion of 
the immersible part of the probe at full length) 
d) Soft, fluff free cloth or single use towel 
e) Personal protective equipment (gloves, water repellent protective skirt, face protection 
mask or protective glasses, see also instructions of the manufacturer for the detergent 
and the disinfectant) 
 
Manual Cleaning:   
Prepare the detergent solution in a tank with cold water (please follow the instructions of 
the detergent manufacturer regarding application, dilution and contact time). 
1) The temperature of the detergent solution should be between 15‐30°C, concentration is 
1.6%.  Please  note  the  minimum  contact  time  of  the  detergent  in  the  manufacturer's 
instruction.  If  a  differing  detergent  is  used,  please  also  note  the  approved  material 
compatibility for the medical device.     
2) Immerse  the  immersible  part  of  the  probe  without  connector 
into  the  diluted  detergent  solution  (see  Fig.  3).  Wipe  the 
immersible part of the probe under the surface of the detergent 
solution with a soft cloth to remove all visible soil. Be sure that 
all  grooves  of  the  probe  are  implemented  during  the  cleaning 
process. 
3) The  immersible  part  of  the  probe  should  be  left  in  the 
detergent  solution  according  to  the  specified  contact  time  of 
the detergent manufacturer. 
4) Wipe  the  un‐immersible  parts  of  the  probe  with  a  soft  cloth 
dipped with the detergent solution. 
5) Rinse  the  probe  with  running  tap  water  for  1  minute. 
(alternatively:  immerse  the  immersible  part  of  the  probe  in  a 
tray  filled  with  deionized  water/tap  water  (see  Fig.  3)  for  5 
min.)   
6) Visually  check  the  outer  surface  of  the  probe  for  cleanness.  If 
necessary,  use  magnifying  glass  for  visually  check.  If  there  is 
still soil visible, repeat all above steps. 
 
 
‐6‐ 
Water, Detergent 
or Disinfectant
Fig. 3  Immersion 
of the probe 
Q1E‐EP1028