CLEANING & CARE – REPOSE ULTRACORE COVER
LIGHT SOILING
! IMPORTANT
In the absence of gross contamination or unusual risk. The removal of dirt and
spillages with a neutral detergent and warm water should be sufficient. However, if required the
cover can be wiped down with a 0.1% hypochlorite solution (1000ppm). Ensure that the cover is
rinsed with clean water using a single use non-abrasive cloth, and thoroughly dried. Ensure
adequate ventilation.
HEAVY SOILING
! IMPORTANT
In cases of gross contamination the mattress cover should be cleaned first with
detergent and water and then with a diluted solution of 0.1% hypochlorite (1,000ppm). Rinse
well with clean water and wipe down with a damp single use non-abrasive cloth, and thoroughly
dried. Ensure adequate ventilation.
! IMPORTANT
If required, a 1% hypochlorite solution (10,000ppm) and cold water may be
applied for no more than 2 minutes.
! IMPORTANT
Machine washing instruction for outer mattress cover.
Disinfection may be achieved by laundering at temperatures not exceeding 65
73
C for 3 minutes. Protect, Rinse and Dry -BHTA guidance on the care, cleaning and inspection
º
of healthcare mattresses. (January 2012 - HHTA HM 0112)
DRYING
! IMPORTANT
To avoid shrinkage, line dry in an indoor clean environment. Covers must be
thoroughly dried before re-fitting to the mattress.
! IMPORTANT
Do not mangle or iron.
INSPECTION
WARNING:
Inspect the inner and outer surfaces of covers and their zip fasteners regularly for signs of
damage. If the cover is stained, soiled or torn, the foam and Repose core should be examined.
Damaged / soiled covers and mattresses should be reported to the ward / department manager.
If the core of the mattress is wet or badly stained, the mattress should be withdrawn from
service.
CLEANING & CARE – REPOSE ULTRACORE OVERLAY
• Repose ultracore overlay can be cleaned with soap or detergent and water and can be
disinfected in accordance with Protect, Rinse and Dry -BHTA guidance on the care, cleaning
and inspection of healthcare mattresses. (January 2012 - HHTA HM 0112)
• Alcohol & phenol based disinfectants should not be used
• Allow to dry before re-use or storage
! IMPORTANT
Disposable gloves and aprons should be worn. As splashing could occur, it is
recommended that eye / face protection should also be worn. On removal of gloves, hands
should be washed. Ensure Repose ultracore overlay is thoroughly dried before placing back
inside the cover.
6
C for 10 minutes or
º