TPI 310C Instructions

Pocket thermometer

Advertisement

310C
Pocket Digital Thermometer
To measure temperature with this unit follow these simple instructions:

Selecting Fahrenheit or Celsius mode:

With the unit "OFF" press and continue to hold the ON/OFF button. The LCD will alternate between °F and °C
Release the ON/OFF button when desired measurement mode appears on the LCD
Remove the protective sheath from the stem.
Press the ON/OFF button.
Insert stem into area or object to be tested. At least 1/2" of the stem must be inserted.
Warning: The stem is a sharp and pointed object which can be dangerous.
The Data Hold function is used by pressing the D-H button when taking a measurement. The measurement will
flash on the LCD when Data Hold is in use. To return to normal function, press and release the D-H button a
second time.
Press the ON/OFF button to end operation, or the unit will automatically power off after 35 minutes.
To reverse the LCD to suit your viewing angle:
Insert a small coin into notch in curved area "B", then gently pry portion "A" from "B". Shielded wires will
keep the two parts connected.
Twist the area "A" 180° and snap back into area "B".

Specifications:

Range
Resolution
Accuracy
Sampling Time
Battery
TPI 310C FIELD CALIBRATION
Fill a plastic or metal container with crushed ice and add clean water to a depth of at least 4 inches. Stir the ice
and water for 2 to 3 minutes prior to performing calibration to ensure the water is completely chilled. Make
certain there is plenty of ice in the mixture and always use clean water. Distilled water works well. The
temperature of an ice bath is approximately 32°F (0°C).
Insert the stainless steel shaft of the 310C into the ice bath making sure at least one inch of the tip is immersed.
Allow the reading on the thermometer to stabilize.
Note: The temperature reading must be within 23°F to 41°F (-5°C to 5°C) for calibration to have effect.
-58° to 300°F/-50° to150°C
0.1°F/C
+2°F/+1°C
1.5 seconds
1.5 Volt Button

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 310C and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for TPI 310C

  • Page 1 32°F (0°C). Insert the stainless steel shaft of the 310C into the ice bath making sure at least one inch of the tip is immersed. Allow the reading on the thermometer to stabilize.
  • Page 2 Press and hold down the HOLD button for approximately 8 seconds until “CAL” is displayed. “CAL” will display for approximately 2 seconds and then the 310C will return to normal operation. Calibration is complete. Note: If the temperature reading was not within 23°F to 41°F (-5°C to 5°C) when step 3 was performed no change in calibration occurred.

Table of Contents