A blank screen will re-appear with a menu bar along the top.
Click on LogTag on the top menu and select „Configure‟ in the drop
down menu.
The program will automatically „Locate LogTag‟ (
Click Next for the Arrow to move down to configure Logtag.
You will now be on Step 2 of the programme > „Configure‟.
Continue through the options and configure the LogTag to your
personal requirements
(Suggestions)
Name your LogTag for identification purposes in the user ID
field.
Disable pre-start logging, unless required.(tick the box- it will
clear)
Set your reading intervals to suit your requirements and
recording duration.
Activate a reading delay (15- 30 minutes) to allow the recorder
to acclimatise, if required
Set your alert parameters to suit your requirements.
The consecutive reading setting "delay", allows for defrost cycles
etc, in an effort not to send "false" alerts for brief excursions.
- After configuration click „Next' – wait for command and click „Next‟
to ensure LogTag is ready for use.
- Wait for the space bar to show 100% to ensure your LogTag has
accepted the configuration. _ You can remove the Logtag NOW.