Right arrow on this menu option to begin the calibration. The
display will read, 'Use fresh air (20.9%) Please wait...'. The F-960
will beep upon completion of the O2 calibration in air.
Setup > Date and Time
To adjust the date and time, use the Right and Left arrows to
move between Month/Day/Year and Hour/Minute/Second and
use the Up and Down arrows to change the values. To exit, use
the Left arrow to back out of the screen and return to the Setup
menu.
Date & Time
MM/DD/YYYY
04/21/2016
Setup > RH Conversion
The Setup RH Conversion menu is used to correct for the
temperature sensor being inside F-960 housing. This causes the
temperature sensor to reflect the temperature of the F-960 and
not the ambient temperature, which is usually several degrees
above ambient. The Relative Humidity is calculated based on
the temperature value.
F-960 Instruction Manual rev. 5/30/2017