6.5.2 Using the preset .......................... 23 6.5.3 Searching for a preset ......................... 24 "Current procedure" application area..................... 25 6.6.1 Patient............................ 25 6.6.2 Checklist ............................ 31 6.6.3 Capture ............................ 33 6.6.4 Edit............................... 38 ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 4
Table 5 – Recommended separation distances between portable and mobile HF communications equipment and the product ........................ 92 Regulatory information on telecommunications .................. 93 9.8.1 USA.............................. 93 9.8.2 Canada ............................ 93 9.8.3 European Union ........................... 93 10 Subsidiaries.............................. 94 ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries. ATTENTION ATTENTION Designates a possibly harmful situation. If this is not avoided, the products could be damaged. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The product does not come into direct contact with the patient. 2.5 Target user populations The medical device may only be used by doctors and medical assistants with a relevant specialist qualification. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
With an OR1 ™ InWall Solution, the temperature and ventilation around the product are continuously monitored. If the temperature around the product exceeds a certain value, an error message is issued. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Do not use any additional multiple socket outlets or extension cables. Do not use freely accessible multiple socket outlets. Multiple socket outlets must not be placed on the floor. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Only start the application when combustible or flammable gases and liquids are no longer present. 3.9 Observing ambient conditions If the product is used in an unsuitable environment, patients, users, and third parties may be injured. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Connecting non-compatible signal types to the patch panel in the OR can cause image interference and malfunctioning. Only connect permitted signal types. Perform equipment tests before use. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Perform a system test which verifies perfect functioning after adding new software or hardware. Shut down the software via the user interface or by briefly pressing the rocker switch. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
OR1 FUSION CONTROL front view Rocker switch ON/OFF "Red" LED for hard disk access Standby and operating mode "Yellow" standby LED USB port "Green" operating LED ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Degree of protection acc. to IEC 60259 IPX0 Housing Dimensions (L x H x W) 355 x 74.5 x 305 mm Weight 6 kg System Hard disk capacity 2 TB Random Access Memory (RAM) 16 GB ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Number of products in the product packaging Unique Device Identifier Consult the printed or electronic instructions for use Fragile, handle with care Keep dry Humidity limit ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Storage/transport conditions Temperature -10 °C ... +60 °C Relative humidity 5–95 % (non-condensing) Air pressure 500–1,080 hPa Operating conditions Temperature 5 °C ... 35 °C Relative humidity 20–80 % (non-condensing) Air pressure 700–1,080 hPa Max. operating altitude 3,000 m ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Nothing may be mounted in the defined free areas. 5.3 Connecting the product The product may only be connected to other components by service personnel or personnel authorized by . ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Last name – First name – Date of birth – – Patient ID Application areas The application areas contain modules that can vary depending on the configuration: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Switch microphone to mute. The microphone is set to mute as standard and must be activated manually. Volume Adjust the output level of the active speakers or headphones. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
A high volume could startle the surgeon. This can result in injury to the patient, surgeon, or third parties. Check the volume before each operation. Volume level Speaker ON/OFF ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
6.5.1 Creating a preset Configure the desired settings. Tap Presets. ð The Preset screen appears: Touch the Create button. ð The following input screen appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 22
Activate the required checkboxes to assign the preset to procedures and physicians. Confirm the entries by pressing OK. ð The preset appears in the preset list: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
ð The Preset screen appears: Touch the desired preset. ð The selection turns orange: Confirm the selection by pressing OK. ð The settings saved in this preset are now set. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The search is limited to the type of procedure. Select the type of procedure. Touch the preset symbol ð The preset screen appears and all presets that contain the search term are displayed: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The administrator can limit or expand the scope of the documentation for the Patient module and this scope needs to be configured in cooperation with the technicians during installation and commissioning. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 26
Enter all patient data. The mandatory fields are blocked once you have captured audio and video recordings. Touch the Modify data button to unlock the mandatory fields. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 27
ð The input screen appears: Enter all patient data. Touch Team in the menu. ð The input screen appears: Enter all the members of the surgical team. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 28
11. The mandatory fields are blocked once you have captured audio and video recordings. Touch the Modify data button to unlock the mandatory fields. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 29
Mark the data set you require and touch the Select patient button. The selected data set appears in the input screen. The imported data set is write-protected as standard and cannot be edited. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 30
ð The input fields are cleared. ð The data stored in the DICOM is preserved. ð The system is ready for the manual compilation of a new procedure. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
ð The selected checklist starts. The selection cannot be canceled. ð Any entered patient and procedure data appears. ð The progress indicator for the editing of the checklist is shown. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 32
ð Checkpoints that have not been documented are indicated in the progress indicator with a warning symbol. Touch the Finish button in the navigation bar to finish the checklist. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
For documentation purposes, the start time can be overwritten with your own term. Left-click to select the capture you want. ð The context menu appears: Touch the Annotations menu item. ð The Annotations window appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 34
ð The start of the capture is confirmed in the list of video sequences. To start another capture from another channel when capturing video, touch the relevant Video button. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 35
Touch the Delete menu item again to remove the deletion mark. 6.6.3.7 Forwarding captures Touch a still, a video sequence, or an audio sequence followed by View. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 36
ð The image, video, or audio ID is displayed on the target screens. The captures appear in chronological order. Touch the side arrows to scroll through the captures. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 37
Operation ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Mark the captures and touch the Delete button to mark the captures for deletion. ð The captures are not saved and are discarded permanently during data storage. Touch the Undo delete button to remove the deletion mark. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 39
ð The still appears on the right next to the preview window. 6.6.4.3 Still editing Stills can be edited in the Still editing tab. Mark a still. ð The still appears in the preview window. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
ð If automatic data storage has been configured, a list of the documentation to be saved appears. ð If destination directories have been configured for manual saving, a selection of destination directories appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Some of the individual modules may not be available, depending on the upgrade level and configuration of the system. The application area contains the following modules: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Intubation is monitored using the Anesthesia Safety module. Routing is not necessary due to the predefined settings. Touch Anesthesia Safety in the main menu. ð The module appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Routing to a target outside the operating room may impinge on the patient's data protection rights. Adhere to the currently valid legislation in the respective country. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 44
The preview image can be displayed in three sizes: – Normal image – Large image – Fullscreen Start the Routing module. ð The normal image appears in the preview: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 45
To switch from the large image to the fullscreen image, touch the screen symbol in the top right. To return from the fullscreen image to the large image, touch anywhere on the image. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 46
Assigning an image source to a target Tap on a source, e.g., Endoskopische Kamera (Endoscopic Camera). ð A preview is shown in the preview area. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 47
ð The signal is shown on the target monitor as a full-screen image. Assigning two image sources to a target Select the desired image composition, for example, PAP in the marked area. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 48
Click in the marked area to assign the source to the same target. ð The signal is shown on the right half of the target monitor. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 49
Select the first source, e.g., Endoskopische Kamera (Endoscopic Camera). ð A preview is shown in the preview area. Click in the marked area to assign the source to a target, e.g., monitor 1. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 50
Click in the marked area to assign the source to the same target. ð The signal is shown on the target monitor as the PIP image. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The field of view of the room camera can be adjusted and saved as a preset. ü The Routing module has been started. Touch the Room camera in the Sources area. ð The operating field for the room camera appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
ð The device is displayed on the touchscreen. Operate the device on the touchscreen. ð The device is synchronized simultaneously. Touch the Home button to return to the screen. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Touch the Previous or Next buttons to play the previous or next song in the list. Activate the Random checkbox to play the songs in the selected category in random order. Activate the Repeat checkbox to repeat the selection infinitely. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Touch a button in the Lighting presets area for approx. 3 seconds to save the settings as a preset. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
– Blinds Touch the desired Room Control module in the main menu. ð The module appears: Touch the desired button. ð The associated function is performed. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
ð The module appears: Touch a button under Select source to import past data to select the desired data source. ð The available data is shown as a table: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 57
Parameter drop-down menu. ð The stills and videos of the selected data set appear: Touch the View button to view the stills and videos. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 58
Activate the copying procedure in Open tasks to complete the copying procedure. If the system is shut down while copying is in progress, the following message appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Touch the Burn CD button to save the data set to a CD or DVD. ð The Label CD/DVD dialog window appears after the burning process. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Operate phone functions. Unified Communication Control external communication. Video Conference Hold video conferences. Room To Room Transmit audio and video data to one or more rooms. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The connections between the rooms are bidirectional. Therefore, in this mode, sources can be forwarded from the user's room and displayed from the other room. The access status Room to Room must be selected in both rooms. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 62
The room OR 1 must have selected the access status Public. – The rooms OR 2, OR 3 and OR 4, etc., can have the access statuses Public, Room to Room or Private selected. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Tap the Streamer entry in the list. ð The StreamConnect ® system interface appears. Further information about using the relevant application can be found in the corresponding instructions for use. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Touch Phone in the navigation bar to call up the phone functions again. 6.9.3.2 Adding a contact Touch the Add button. ð The Add contact dialog window appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 65
Confirm the entry by pressing OK. 6.9.3.4 Deleting a contact Touch the entry to be deleted and then touch the Delete button. ð The entry is deleted. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The preview of the is not to be used for diagnostic or therapeutic purposes. The preview image can be displayed in three sizes: – Normal image – Large image – Fullscreen ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 67
Touch the OK button to return from the large image to the normal image. To switch from the large image to the fullscreen image, touch the screen symbol in the top right. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 68
Contacts and virtual rooms for a video conference can be managed in the Contacts area. ü The Unified Communication module has been started. Touch the input field in the Contacts area. ð The onscreen keyboard appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 69
Touch the Delete button. ð The contact or virtual room is removed from the contact list. 6.9.4.5 Calling contacts ü The Unified Communication module has been started. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 70
ð The following window is displayed to the sender: 6.9.4.6 Selecting the source ü The Unified Communication module has been started. Touch the Select Source button. ð The following window appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 71
Touch the Select Target button. ð The following window appears: Use check marks to select targets. ð The targets turn orange: Confirm the selection by pressing OK. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 72
Touch the deactivated source for at least 3 seconds to activate a video. ð The symbol and the name of the source are displayed as normal. ð The blue standby symbol disappears. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The Video Conference module provides an interface for video conferences in which audio and video data is transmitted, e.g., from endoscopic cameras and room cameras. Touch Video Conference in the main menu. ð The module appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 74
6.9.5.1 Contact list All possible video conference partners are displayed and saved via the contact list. Only the contacts saved in the system can be displayed here. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 75
Enter a search term and confirm by pressing Return. ð The possible contacts are displayed. Tap the desired contact. ð The selection turns orange. Tap Anrufen (Call). ð The connection is established. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 76
ð The source is available to the recipient and appears in the Shared Source area: 6.9.5.4 Selecting the target ü The Unified Communication module has been started. Touch the Select Target button. ð The following window appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 77
Touch the deactivated source for at least 3 seconds to activate a video. ð The symbol and the name of the source are displayed as normal. ð The blue standby symbol disappears. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 78
ð The button turns orange and the video conference ends. ð The following message appears if the access status is changed during an active video conference: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Via the "Room To Room" access status: Connection between 2 rooms (communication in both directions). Touch Room To Room in the main menu. ð The module appears: Room selection Source and target selection Preview ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 80
ð The message in operating room 1 disappears. 6.9.6.2 Establishing a room-to-room connection ü The Room To Room module has been started. Touch the desired room in the Select Room area. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 81
ð The Room To Room symbol flashes in the receiver navigation bar after confirmation. Touch the Room To Room symbol in the receiver navigation bar. ð The Room To Room module appears: ð The following window appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 82
ð The source is available to the recipient and appears in the Shared Source area: 6.9.6.4 Selecting the target ü The Unified Communication module has been started. Touch the Select Target button. ð The following window appears: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 83
Once a target has been selected and confirmed by pressing OK, the signal only appears at the selected targets depending on the upgrade level and configuration of the system. The preview on the touchscreen is no longer available. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 84
ð The connection is disconnected. ð The room you were connected to is informed of the change in status. ð The selected sources disappear from the Shared Local Sources area. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
ð The software will shut down and the product will switch to standby mode. Alternatively, briefly press the rocker switch. ð The software will shut down and the product will switch to standby mode. Switch off all connected devices. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Contact , a branch or an authorized dealer to find out the address of the collection point in your area. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Touch monitor Smart screen 0.2 / 3.0 Smart screen HDMI LAN (2x) USB (front) USB 2.0 (back, 4x) Keyboard, mouse USB 3.0 (back, 4x) PS/2 Keyboard, mouse ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 89
315° phase angles ated with an uninter- Voltage interruption: Voltage interruption: ruptible power supply 100% for 250/300 cy- 100% for 250/300 cy- or a battery. cles cles ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 91
= 1.2 √P 80 MHz to 800 MHz d = 2.3 √P 800 MHz to 2.5 GHz Interferences may occur in the vicinity of equipment marked with the following sym- bol: ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
Page 92
HF communications equipment (transmitters) and the product as recommended below, according to the output energy of the communications equipment. ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...
The device must be able to withstand harmful interference, including interference which may lead to undesirable functions. Includes IC: 1000M-8265NG 9.8.3 European Union The device may only be used indoors. Radio Equipment Directive (RED) 2014/53/EU ® Instructions for use • KARL STORZ OR1 FUSION , Rel. 1.4.2 • JEB924_EN_V1.0_02-2021_IFU_CE-MDR...