Cleaning and Disinfecting ergoline Medical Devices 201000641000 • Version 2020-10-05 / Rev 00 • English...
Page 3
Medical Devices 201000641000 • Version 2020-10-05 / Rev 00 • English This document is a supplement to the Operator‘s Manual supplied with each ergoline product. The complete device documentation includes both the Operator‘s Manual and this document. The most recent version of all Operator Manuals for specific products are available for down- load at the ergoline website www.ergoline.com.
Page 4
Cleaning and Disinfecting ergoline Medical Devices...
Only a soft, dry or moistened cloth with one of the cleaning agents listed in section 1.1.1 should be used for cleaning. The parts should be disinfected after each use with one of the disinfectants listed in section 1.2.1. Cleaning and Disinfecting ergoline Medical Devices...
• Cuff, microphone and tubing may not under or mild soap water. any circumstances: • be immersed in liquids, Allow the microphone to dry before reinserting it in its • be cleaned in a water bath or in running water. pocket. Cleaning and Disinfecting ergoline Medical Devices...
Figure 2 – 2: Inserting the microphone 2.6.5 Inspection and Functional Test Visually inspect the cuff and tubing for damage. Worn, porous, or otherwise damaged cuffs and tubing must not be used. Cleaning and Disinfecting ergoline Medical Devices...
• Spray the disinfectant on the inside of the electrodes. Observe the exposure time specified by the manufacturer. • Disinfect the sealing lips and electrode surfaces with disinfection wipes or impregnated cleaning tissues. Cleaning and Disinfecting ergoline Medical Devices...
Manual cleaning and disinfection Perfektan TB Dr. Schumacher GmbH Manual disinfection www.schumacher-online.com Descoton forte (aldehyde) Manual disinfection Neodisher LM2 Chem. Fabrik Dr. Weigert GmbH & Co. KG Manual cleaning www.drweigert.de Validated alkaline cleaning agents and disinfectants Cleaning and Disinfecting ergoline Medical Devices...
• Wipe the cables with a moist cloth (soap water) – never immerse the cables in liquids. 2.8.4 Inspection and Functional Test Visually inspect the sensors for damage. Worn, deformed, porous, or otherwise damaged sensors must not be used. Cleaning and Disinfecting ergoline Medical Devices...
2 Cleaning and Disinfecting the Device Components 2.9 Ear Clip • Wipe the ear clip with a cloth moistened with 70% isopropyl alcohol. • After cleaning, wipe the ear clip with a clean cloth moistened with water. Cleaning and Disinfecting ergoline Medical Devices...
Page 16
201000641000 • Version 2020-10-05 / Rev 00 • English Cleaning and Disinfecting ergoline Medical Devices...