Operating Environmental Conditions; Operating Procedure - Pintek DP-16V Instruction Manual

High voltage
Hide thumbs Also See for DP-16V:
Table of Contents

Advertisement

4. OPERATING ENVIRONMENTAL CONDITIONS

Temperature
Relative Humidity
(1) Dimensions and Weight:
218 x 83 x 30 mm; 390g
(2) Electrical Safety to IEC 1010-1
! Dual Insulation
! Installation Category III
! Degree of Pollution 2
! Rated Voltage or Max Live-Earth: 8KV RMS
(3) CE Mark
Conforms to EN 50081-1 and 50082-1 standards

5. OPERATING PROCEDURE

! Connect to BP-366N and BP-266N(or BP-276N) to DP-16V input
and place BP-266N(BP-276N) on the circuit to be tested.
! Connect BP-250 and DP-16V output and connect to oscilloscope.
! Adjust the vertical zero adjustment of the oscilloscope if
necessary.
! Select the attenuation ratio* and the vertical deviation of the
oscilloscope in accordance with the conversion table below.
! NB: The POWER light must come on.
The conversion table gives the real vertical deviation.
Attenuation
X 2000
X 200
X 20
Reference
+20°C ... +30°C
70 % RH
MAX Voltage Input Range
(DC+AC Peak)
16KVp-p
1.6KVp-p
160Vp-p
Use
0°C ... +50°C
10 % ... 85 % RH
AC RMS
MAX Input
5.6KVrms
560Vrms
56Vrms
4
DP-16V Instruction Manual
Storage
-30°C ... +70°C
10 % ... 90 % RH
DCV
MAX Input
± 8KV
± 800V
± 80V