DP-16V HIGH VOLTAGE DIFFERENTIAL PROBE

1. FEATURES

The DP-16V differential voltage probe provides a safety means of
measuring floating potentials for all models of oscilloscopes
incomplete safety.
It converts the high differential voltage (
voltage (
oscilloscopes. It is designed specially for high voltage using.
Frequency is up to 100MHz.
The BNC output is designed to operate on an input with an
impedance of 1 MΩ. It is 2 times of the 50Ω.
DP-16V can observe more exact measured testing voltage from
DMM. (Oscilloscope accuracy is 3%, and DMM is 10 times).

2. SPECIFICATIONS

(1) Bandwidth:
DC to 100 MHz (-3 dB)
(2) Attenuation: x 10, x 100, x1000
(3) Accuracy: ± 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
16KV for x 2000, (i.e about 5.6KV RMS or ±8KV DC)
1.6KV for x 200,
160V for x 20,
(5) Permitted Max Input Voltage
Max differential voltage: 16KV (DC + AC peak to peak)
Max voltage between each input terminal and ground:
8KV RMS
(6) Input Impedance:
Differential: 30M
Between terminals and ground: 15M
4.0V) with reference to the earth for display on the
(i.e about 560V RMS or ±800V DC)
Ω
// 1.5 PF
(i.e about 56V RMS or ±80V DC)
Ω
// 3 PF
2
16KV peak) into a low