Troubleshooting Chart - Idex Gast DOA Series Operation & Maintenance Manual

Laboratory oil-less diaphragm vacuum pump & compressor
Table of Contents

Advertisement

PART NO. 70 - 5105 G471PL (Rev A)

TROUBLESHOOTING CHART

Low
Vacuum
Pressure
8
High
Pump
Vacuum
Pressure
Overheat
At Pump
At pump
At pump
At pump
At pump
ISO 9001 & 14001 CERTIFIED
Motor
Excess
Overload
Noise
inc.
Reason and remedy
for problem.
Filter dirty. Clean or replace.
Muffler dirty. Clean or replace.
Valves dirty or valves bent.
Clean or replace.
Worn diaphragm.
Repair or replace.
Relief valve set too high.
Inspect and adjust.
Relief valve set too low.
Inspect and adjust.
Plugged vacuum/pressure line.
Inspect and repair.
Collapsed vacuum line.
Inspect and repair.
Low voltage, won't start.
Check power source.
Voltage wrong.
Check power source.
Worn diaphragm.
Replace.
Leaky hose or check valve.
Replace.
Dirt or liquid on top of diaphragm.
Inspect and clean.
Motor not wired correctly.
Check wiring diagram/line voltage.