FREEZER COMPONENTS
1. Lock: Turn clockwise to 180
2. Magnetic door gasket: This prevents the cold air from escaping. Always keep clean.
3. Handle: Always grip this handle to open and close the door.
4. Leveling foot: The height of the freezer can be adjusted by this screw type foot. Keep the unit in
level at the installation.
5. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling foot so that the front two casters cannot contact with the floor.
6. Shelf: Items to be stored in the chamber must be placed on the shelves. The maximum storage
weight for each shelf is 50 kg.
Note: Fix the shelf stopper and shelves securely. Incomplete installation may cause injury or damage.
Never touch the storage items with wet hands. Touching with the wet hands may cause frostbite
7. Access port: This is used for leading the measuring cable from the chamber to the outside. A total of
two (2) are provided. ; left side and top of the freezer.
8. Door latch: Turn clockwise to lock the door and also add the lock too.
9. Temperature sensor: Located on the rear of the chamber.
10. Cold air exhaust vent: Ensure this vent is never blocked. If this vent is blocked, temperature
regulation become unstable. Do not place stored item in the path of cold air.
11. Air intake vent:
temperature distribution in the chamber.
12. Control panel: The operation status is displayed on this panel. And the temperature setting is
available through this panel. Refer to page 10 for details.
13. Front cover: Open this cover when connecting the remote alarm or using the power supply outlet.
14. Space for temperature recorder:
available separately. See page 26 "TEMPERATURE RECORDER (OPTION)".
15. Battery switch: Switch for battery used for power failure alarm. Always keep ON. Turn the
switch OFF when the unit is in no use for a long period (more than one month).
16. Remote alarm terminal: Used to notify an alarm condition of the unit to remote location. See
page 14 for details.
17. Evaporating tray (back side): Defrost water from the evaporator accumulates in the evaporating
tray and evaporates into the atmosphere. See page 17 for cleaning.
18. Power switch (back side): This is for turning ON/OFF the power to the unit. ON – "I" OFF – "○"
19. Fixture (back side): These keep the adequate apace between the unit and wall, and also can be
used for fixing the unit.
o
with a key and the door is securely locked.
Ensure this vent is never blocked.
Space for a temperature recorder (optional component) is
Failure to do so will result in unstable
9