output moDe

The M9.06 pH/ORP monitor and transmitter features 2 solid state relays and 2
mechanical relays in addition to 2 analog output 4-20mA.
Only the second mechanical relay can be set as an alarm (icon is 4ALR)
related to the feedback of external device managing. Icon will turn to 4OTA
(Over Time Alarm) in case the setpoint has not been reached within set
maximum timing.
Icon will turn to 4OVA (Over Values Alarm) in case measured values overstep
the set value band. In addition to the type of failure, a reference number
correlated to the involved digital output is reported by the out put number.
PROCEDURE FOR OUTPUTS SETTING
- go to the "Options" menu
- enter into the "Outputs activation" sub menu
- enable output(s)
- go to the "Outputs" menu
- set the operating mode for each enabled output
ENTER TO CONFIGURE
Monitor without digital
output activated
12
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
1 OFF
pH
ENTER TO CONFIGURE
In case a digital output
is enabled, a icon will
appear
1 PRP
1 PRP
1 PRP
1 PRP
pH
ENTER TO CONFIGURE
In case a digital output
is set, icon reports the
operating mode
1 PRP
pH
ENTER TO CONFIGURE
In case set digital
output is activated, the
icon will turn to black
pH