Page 1
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 1 of 16 by S Instruments, Inc. Patient Floor Scale Operating and Service Manual SInstruments, Inc. 600 Young Street, Tonawanda, NY 14150 Tel: 716-693-5977 Fax: 716-693-5854 URL: www.srscales.com...
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 2 of 16 TABLE OF CONTENTS PACKING CHECKLIST – MODEL SR416I-KG ................3 ASSEMBLY............................3 SYSTEM DESCRIPTION AND INTENDED USE ...............3 STORAGE AND TRANSPORTATION ..................4 MICROBAN ANTIMICROBIAL PROTECTION ...............4 CLEANING AND DISINFECTING ....................5...
Service Hotline at: 1-800-654-6360. STEP 2: Place the SR416i-KG scale system on a flat, hard floor surface. Ensure all four corners of scale are in contact with the floor and scale does not rock.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 4 of 16 STORAGE and TRANSPORTATION To maintain proper operation of this instrument, storage and transport conditions should not vary outside the following conditions: Relative Humidity Range of 0% to 85%, Ambient Temperature Range of 14F to 122F (-10C to 50C).
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 5 of 16 CLEANING and DISINFECTING CLEANING To clean the display / user interface and other scale contact areas: Use a soft cloth dampened with water and mild detergent to clean scale surfaces.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 6 of 16 SPECIFICATIONS MAXIMUM WEIGHT 400 kg CAPACITY 26.6 in x 18.5 in (68 cm x 47 cm) DIMENSIONS DISPLAY TYPE 1” Digit LCD DISPLAY RESOLUTION +/- 0.1 kg...
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 7 of 16 BUTTON FUNCTIONS Figure 1: SR416i-KG Button Label ZERO / WEIGH The “Zero/Weigh” button is used to zero the system before placing a patient onto the scale system.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 8 of 16 BASIC SYSTEM OPERATION PATIENT WEIGHING STEP 1: Place scale on hard smooth floor surface to ensure accurate data. NOTE: Avoid carpeting or uneven flooring surfaces.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 9 of 16 THEORY OF OPERATION SR Instruments patient weighing systems are digital scales. Strain-gauge force cells convert the force of an applied weight into an analog signal. This signal is amplified by an operational amplifier and converted to a digital signal by analog to digital converter.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 10 of 16 CALIBRATION IMPORTANT CALIBRATION CHECK: Qualified service personnel only should perform this procedure. Load cells have no user serviceable components and should not be tampered with for any reason.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 11 of 16 CALIBRATION (Cont’d) ACCESS SYSTEM SETUP The Right Button allows you to toggle through the menus without saving any parameters. The Left Button permits a menu item to be selected or a parameter saved (Figure 3).
Page 12
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 12 of 16 CALIBRATION (Cont’d) Auto Power Down Timer: (Optional) a) Select the Auto Power Down Timer from the CAL menu by pressing the Left Button when the display shows “sec”.
Page 13
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 13 of 16 CALIBRATION (Cont’d) Half-Scale value: (Optional) NOTE: The Half-Scale value is used only with the three-point calibration. It is a value between zero and the Full Scale values. It is usually close to half of the Full Scale value.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 14 of 16 TROUBLESHOOTING SYMPTOM REASON/CORRECTIVE ACTION Weight reading is much lower than Check that the platform is clear underneath, stand on each expected. corner to see if one corner is not weighing correctly.
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 15 of 16 WARRANTY FOUR (4) YEAR LIMITED WARRANTY Each S system is manufactured with high quality components. SR Instruments, Inc. warrants that all new equipment will be free from defects in material or workmanship, under normal use and service, for a period of four (4) years from the date of purchase by the original purchaser.
Page 16
Model SR416i-KG - Patient Floor Scale Operating and Service Manual Part No. MAN416I-KG_200611 Page 16 of 16 By SInstruments, Inc. Precision & Technology in Perfect Balance SInstruments, Inc. 600 Young Street, Tonawanda, NY 14150 Tel: 716-693-5977 Fax: 716-693-5854 URL: www.srscales.com...
Need help?
Do you have a question about the SR416i-KG and is the answer not in the manual?
Questions and answers