VWR International 2310 Installation And Operation Manual

Water jacketed co2 incubators with microprocessor control

Advertisement

sales@artisantg.com
artisantg.com
(217) 352-9330 |
|
Click HERE
Find the VWR / Shel Lab 2300-MP at our website:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 2310 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for VWR International 2310

  • Page 1 sales@artisantg.com artisantg.com (217) 352-9330 | Click HERE Find the VWR / Shel Lab 2300-MP at our website:...
  • Page 2 WATER JACKETED CO2 INCUBATORS WITH MICROPROCESSOR CONTROL MODEL: 2310, 2300, 2350 2400, 2450 INSTALLATION AND OPERATION MANUAL Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@Shellab.com INTERNET: http://www.Shellab.com/~Shellab 1-800-322-4897 (503) 640-3000 FAX (503) 640-1366 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 3: Table Of Contents

    TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION SECTION 4.0 CONTROL PANEL OVERVIEW SECTION 5.0 OPERATION SECTION 6.0 FYRITE CO2 CHECKING SECTION 7.0 MAINTENANCE SECTION 8.0 TROUBLESHOOTING SECTION 9.0 PARTS LIST UNIT SPECIFICATION SCHEMATICS REV 02/07 4861519...
  • Page 4 INTRODUCTION Thank you for choosing a water jacket incubator. These units are not intended for use at hazardous or household locations. Before you use the unit, read this entire manual carefully to understand how to install, operate, and maintain the unit in a safe manner. Your satisfaction with the unit will be maximized as you read about its safety and operational features.
  • Page 5: Receiving And Inspection

    Carefully check all packaging before discarding. accessory package is included with each unit. The 2310 comes equipped with 3 shelves, 4 shelf standards, 4 leveling feet, 1 humidity pan, and supply hose kit. The 2300 and 2400 comes equipped with 3 shelves, 2 shelf stands and 1 humidity pan per chamber, 4 leveling feet and 1 supply hose kit.
  • Page 6: Graphic Symbols

    Section GRAPHIC SYMBOLS Your incubator has been provided with a display of graphic symbols, which are designed to help in identifying the use and function of the available user adjustable components. Indicates that you should consult your manual for further description or discussion of a control user item.
  • Page 7: Installation

    Section INSTALLATION Local city, county, or other ordinances may govern the use of this equipment. If you have any questions about local requirements, please contact the appropriate local agency. Installation may be performed by the end user. Under normal circumstances this unit is intended for use indoors, at room temperatures between 5°...
  • Page 8 Leveling: The unit must sit level and solidly. Leveling feet are supplied and must be installed in the four holes in the bottom corners of the unit. With the feet installed and the unit standing upright, each foot can be raised by turning it in a counterclockwise direction.
  • Page 9 mounting grooves for the door gaskets. Remove all shelves, shelf supports, shelf standards and shields. Autoclave, or wash and disinfect as described in item 1. Wash and disinfect all interior surfaces. Give special attention to cleaning and disinfecting all access ports, air bleeds, shaft holes, electrical pass-through and any other passages into the chamber.
  • Page 10 Figure 1 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 11: Control Panel Overview

    Section CONTROL PANEL OVERVIEW (See Figure 2) Units with a detachable cord have a fused inlet located at the top rear of the control panel. This inlet has a recessed male plug, fuse, and an EMI filtering system designed to filter out electrical interference.
  • Page 12 injected. The TC type CO2 sensor required 95% humidity to correctly measure the CO2 levels. The CO2 should be turned OFF during warm up. Failure to do so will cause the control overshoot the set value input by the operator, the injection value can be reset by going into cal mode on the main and door temperatures and back out of cal mode.
  • Page 13: Operation

    FILL WJ HERE. Stop filling when the WATER LOW warning light goes off. When full, the 2310 holds approximately 9.5 U.S. gallons (35.96 Liters) and 19.0 U.S. gallons (72 Liters) for the 2300/2400 chambers.
  • Page 14 Setting the Controls: Refer to the Control Panel Overview Section 4.6 of this manual. Calibrating the Controls: Push and hold both the UP and DOWN buttons until the decimal points in the display begin to blink. While the decimal points are blinking the UP and DOWN buttons can be used to adjust the display to match the actual condition in the incubator chamber.
  • Page 15 All units are shipped set for 37°C and 0.0% CO2. If other conditions are desired follow the instructions in section 5.3 for resetting controls. Temperature Calibration: It is important that the calibration be performed in the following order: Main temperature, then the over temperature safety. During main temperature calibration it is important that the inner glass door not be opened for any reason, and that the outer heated door be opened as little as possible.
  • Page 16 Figure 3 9850677 The CO2 inlet fitting is located on the back of the incubator near the top (see figure 3). It is marked "CO2 TO CHAMBER". The supply hose with in-line CO2 filter (supplied with your accessories) connects from this fitting to the CO2 tank and regulator.
  • Page 17 There are two types of CO2 sensors available, Thermal Conductivity (TC) and Infrared (IR). The 2310 and 2300 units operate with TC sensors, and the 2400 units operate with IR sensors. TC sensors operate under the principle that CO2 is 1.5 times denser than air and if you measure the changes in density they can be interpreted as changes in CO2 content.
  • Page 18 CO2 set point, measure the actual CO2% with a Fyrite. If there is any difference between the Fyrite and the display, use the procedure described in section 5.4 for making the display match the Fyrite. See Section 6.0 for Fyrite use. NOTE: When using the Fyrite, insure that gas is not being injected while the reading is being taken.
  • Page 19 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 20: Fyrite Co2 Checking

    Section FYRITE CO2 CHECKING A Bacharach FYRITE CO2 Gas Analyzer is recommended to measure CO2 concentrations in the incubator chamber. This test instrument is not supplied with the incubator but is readily available from your dealer. Follow the instructions provided with the Fyrite instrument carefully to insure correct and accurate readings.
  • Page 21: Maintenance

    Section MAINTENANCE NOTE: Prior to any maintenance or service on this unit, disconnect the service cord from the power supply. Cleaning: Cleaning and decontamination are recommended on a regular basis. To prepare the incubator for cleaning remove all parts such as the shelves, shelf standards and false ceiling.
  • Page 22: Troubleshooting

    Section TROUBLESHOOTING CAUTION: Extreme caution must be exercised any time access is made into areas housing electrical components. Repair, replacement or adjustment of components in these areas must only be done by qualified technicians familiar with electrical circuitry and the operation of the chamber. Always make a visual inspection of the incubator and control console when troubleshooting.
  • Page 23 9/ loose connection – check shadow box for loose connections Display reads "LO" 1/ sensor is plugged in backwards – reverse sensor wires to controller 2/ if ambient room temperature is lower than range of unit – compare set points and ambient temperature to rated specifications in section 9.0 Unit will not heat over a temperature that is below set point...
  • Page 24 temperature 2/ see if ambient is fluctuating Display and reference thermometer don’t match 1/ calibration error – see section 5.4 2/ temperature sensor failure – evaluate if pilot light is operating correctly 3/ controller failure – evaluate if pilot light is operating correctly 4/ allow at least two hours to stabilize 5/ verify that reference thermometer is certified Can't adjust set points or calibration...
  • Page 25 2/ solenoid failed while closed 3/ CO2 controller output failed while open 4/ restrictor plugged 5/ CO2 hose blockage 6/ CO2 filter plugged 7/ set point is at 0.0 and has not been reset, see section 5.3 and Display and Fyrite reading do not match 1/ calibration error - clear chamber for 12 hrs and confirm at "0"...
  • Page 26 2/ check output signal from controller 3/ check solenoid valve for correct operation 4/ leak in plumbing including between regulator and CO2 tank 5/ door being opened too often Won’t hold calibration – Fyrite reading varies but display stable 1/ atmospheric pressure fluctuations 2/ top of unit exposed to cold air drafts 3/ unit being operated without shadow box cover in place 4/ condensation collecting on CO2 sensor...
  • Page 27 Unit or wall fuse/circuit breaker is blown 1/ check wall power source 2/ compare current draw and compare to specs on data plate 3/ see what other loads are on the wall circuit Unit will not turn on 1/ check wall power source 2/ check fuse/circuit breaker on unit or in wall 3/ see if unit is on, e.g., fan or heater, and just controller is off 4/ check all wiring connections, esp.
  • Page 28: Parts List

    Section PARTS LIST Description 115V 220V Anode 0260500 0260500 CO2 Display 1750559 1750560 CO2 Control w/Alarm – 2310 only 1750551 1750552 CO2 Filter 100199 100199 Control 4 Loop Pad 1750725 1750725 Cord Set 1800510 104192 Door Switch X1000022 X1000022 Door Heater...
  • Page 29: Unit Specification

    UNIT SPECIFICATIONS These units are 115 volt or 220 volt. Please refer to the unit data plate for its individual specification. Weight Shipping 2300 280 lbs. 221 lbs. 2310 280 lbs. 257 lbs. 2350 600 lbs. 442 lbs. 2400 280 lbs.
  • Page 30 Model Voltage Hertz 2300-2 3.5 amp 220V 50/60 Hz 2310 5 amp 120V 60 Hz 2310-2 3.5 amp 220V 50/60 Hz 2350 5 amp 120V 60 Hz 2350-2 3.5 amp 220V 50/60 Hz 2400 5 amp 120V 60 Hz 2400-2 3.5 amp...
  • Page 31 WIRING DIAGRAM 2300,2310,2350 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 32 WIRING DIAGRAM 2400 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 33 SHELDON MANUFACTURING, INC. LIMITED WARRANTY Sheldon Manufacturing, Inc., (“Manufacturer”) warrants for the original user of this product in the U.S.A. only that this product (parts only if outside of the U.S.A.) will be free from defects in material and workmanship for a period of two years from the date of delivery of this product to the original user (the “Warranty Period”).
  • Page 34 Bridgeport, NJ San Diego, CA Lake Charles, LA S. Plainfield, NJ San Dimas, CA Or Call Direct for Specialized Service Locations: VWR International Switzerland VWR Furniture Division 3000 Hadley Rd Ruchligstrasse # 20 P.O. Box 3405 S. Plainfield, NJ 07080 P.O.