9. Sampler set-up -
time-based sampling with the Optional DPS-2
9.1 Calculate the sampling rate using the following 30 day chart:
Number of turns
sample
open on pump
pump
stroke knob
displacement
per stroke
3
6
9
12
9.2 Set the timer dials on the Z-65 to the
determined time from step 9.1.
NOTE: To obtain maximum battery life, choose
the longest time interval and the largest pump
displacement setting possible.
Sample cylinder volumes
1000 cc
.100
4
.200
9
.300
13
.400
18
300 cc
500 cc
9
15
18
30
27
45
36
60
Example
18 minutes
0
NOTE: The time (18 minutes) above
corresponds to the dial setting shown for the Z65
model with the timer range setting in the factory
position (jumper on the two left pins). See section
12.4 Timer Range Setting.
Sample
rate
(minutes)
0