6.2 Managing sets of measuring parameter
In the alphaDUR mini combinations of measuring parameter can be stored,
under a user-defined name. In this way, the measuring parameter required for
specific applications can be called up easily.
The following parameter are stored:
Name
Test load
Material
Hardness scale
Upper and lower limit for the rating ´GOOD´
Test report printout setting (on or off)
The number of values which should be used for the statistical
evaluation
6.2.1 Editing measuring parameter
Under the menu item Measuring parameter / Edit, the measuring parameters
can be set.
Just the current measuring parameters can be changed. In order to change a
stored configuration, it must first be loaded and stored again after the changes
has been implemented.
6.2.2 Saving measuring parameter
The current measuring parameters can be stored in the menu item Measuring
parameter / Save under a user-defined name.
After selecting the menu item, a text entry window is opened in which the new
name for the set of measuring parameter can be entered.
MEASURING PARAMETER
21