HBP-1300-E main.book Page 33 Thursday, February 11, 2016 11:39 AM

Manufacturer's Declaration

The HBP-1300 is intended for use in the electromagnetic environment specified below.
The customer or the user of the HBP-1300 should assure that it is used in such an environment.
Electromagnetic Emissions:
(IEC60601-1-2)
Emission Test
RF emission
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
IEC 61000-3-3
Electromagnetic Immunity:
(IEC60601-1-2)
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electric fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
The HBP-1300 uses RF energy only for internal functions.
Therefore, this RF emission is extremely weak and there is
little chance of it creating any kind of interference
whatsoever with nearby electronic equipment.
Class B
The HBP-1300 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
Class A
that supplies buildings used for domestic purposes.
Complies
IEC60601-1-2
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for
±2 kV for
power supply lines
power supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV
±1 kV
differential mode
differential mode
±2 kV
±2 kV
common mode
common mode
<5 % U
<5 % U
T
T
for 0.5 cycle
for 0.5 cycle
40 % U
40 % U
T
T
for 5 cycles
for 5 cycles
70 % U
70 % U
T
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
T
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Electromagnetic Environment
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
HBP-1300 requires continued
operation during power mains
interruptions, it is recommended that
the HBP-1300 be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
EN
33