GE Dinamap PRO Series Manual page 40

Hide thumbs Also See for Dinamap PRO Series:
Table of Contents

Advertisement

Step
Description
Measured temperature in °F (80.2 nominal)
5.3.4.5
Measured temperature in °F (107.8 nominal)
5.3.4.7
5.3.4.10 Broken Probe fault condition
5.4
Pushbutton test
5.4.1.1
Verify appropriate responses to pushbuttons
5.4.2
Sound test
5.4.2.2
Verify volume levels
5.4.3
Printer test
5.4.3.2
Printer produces a clear printout
5.4.4
Battery System test
5.4.4.1
Verify AC mains indicator
5.4.4.4
Verify uninterrupted battery operation
5.4.4.5
Verify battery indicator
5.4.5.
Remote Alarm
5.4.5.2
Voltage between pins 4 and 8, alarm inactive
5.4.5.4
Voltage between pins 4 and 8, alarm active
5.4.6
Communications test
5.4.6.2
Verify pump starts, stops and temp. status returns
6.0
Hipot Tests
6.1
AC Main Hipot
6.2
Temp Hipot
6.3
SP02 Hipot
7.0
Safety Tests
7.2.7
EXTDC to GND Resistance (mΩ)
7.3.6
SPO2 leakage (uA) (models 300 and 400)
7.4.6
Temp leakage (uA) (models 200 and 400)
7.5.4.1
Normal no-fault leakage (uA)
7.5.4.3
Normal open-ground leakage (uA)
7.5.4.6
Normal open-neutral leakage (uA)
7.5.5.1
Reverse no-fault leakage (uA)
7.5.5.3
Reverse open-ground leakage (uA)
7.5.5.6
Reverse open-neutral leakage (uA)
Tested By: ______________________________________
A-4
Appendix A
DINAMAP® PRO Series 100-400V2 Monitor Service Manual
2018554-001
Min
Max
Actual Pass Fail N/A
79.9
80.5
107.5 108.1
4.7
5.3
0.0
0.1
0
1000
0
150
0
50
0
500
0
500
0
500
0
500
0
500
0
500
Date: ___________
Revision A