PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they
will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb. The electrode cannot
function properly under these conditions. These bubbles can be removed by "shaking down" the
electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in
HI70300
Storage Solution for at least one
hour.
For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1") below the fill hole add
HI7082
or
HI8082
3.5M KCl Electrolyte Solution for double junction electrodes.