Span Calibration - Option 3 Configuration
1. Fill a gas bag with target sample gas
2. Enter "Calibration" menu of SQN8x for the desired channel
3. Attach the tubing to the desired inlet point of the SQN8x
4. Attach SQN8x to gas bag port.
The sample pump will draw the sample gas from the gas bag
Regardless of gassing configuration, the SQN8x must be in "Calib Span" mode to span the unit.
NOTE:
0ERFORM A :ERO #ALIBRATION BEFORE ATTEMPTING A 3PAN #ALIBRATION %NSURE THE BACKGROUND IS FREE OF POTENTIAL
gases or apply zero grade air
1. Navigate to Programming > Calib Span.
2. Select the sensor to be zeroed.
SQN8x
Calibration Handbook
Span Gas
Sample Gas Bag
Sample Gas Bag
PROGRAMMING
CALIB SPAN
CALIB SPAN
T1
O2
131
1LPM
Regulator