WatchGas BAT-1, 2, 3 Breathing Air Quality Test Kit
Full Compliance with BS EN12021
Figure 2: 200 & 300 bar compressor adapters
1.
Clean the valve of compressed gas cylinders or char-
ging connections of the compressor with a brief burst of
air.
Figure 4: Pressure reducer

5.2 MEASURING ON COMPRESSED GAS CYLINDERS

(for measuring on compressors go to 5.3)
1.
Slowly open cylinder valve.
Figure 6: Checking the outlet flow with the flow meter
2.
The flow rate of air through each port of manifold
can be confirmed if necessary by using the flow meter
provided with the kit. (To measure the flow rate connect
tubing of flow meter to the desired manifold port- Fig. 6)
User Manual
Figure 3: Screwing the adapter on the pressure
2.
Chose and connect the adapter suitable for the pre-
script maximum pressure of the compressor charging
panel (300 or 200 bars). When using the BAT kit on a gas
cylinder, you can skip this step.
Figure 5: Connecting manifold to the pressure reducer
3.
Insert and firmly fit the nozzle of manifold on the pres-
sure reducer as shown in fig. 5.
Figure 7: Connecting rubber tube holders to the manifold
3.
Insert the detector tube holders in the manifold ports-
Fig. 7.
4.
Flush device for sufficient time to make the test device
free from other gases if any.
9
D e t e c t i o n
D e t e c t i o n
reducer.
www.watchgas.com