Operation Manuals High-Speed Refrigerated Centrifuge Model: GL10-MC Thanks for usingDRAWELL centrifuge! In order to get quick and sincere service, please read the operation manuals carefully before you operate.
Contents Contents........................... 1 Attention........................... 1 Attention for safety..........................1 Attention for fireproofing.........................2 Attention for safe operation........................2 Attention for safety in chemistry and biology..................3 1. General introduction....................5 1.1 Application and scope........................5 1.2 Name and Classify........................... 5 1.3. Centrifugation Theory........................5 2.
Attention Thanks for usingDRAWELL centrifuge, to prevent any potential accident,please operate the centrifuge accortding to the following safty operation. Attention for safety Attention for safety clarifies the requirements of safe operation in the instruction manual. Please carefully read it before installation, operation, maintenance and repairing.Understanding the warning and correct operation can avoid the personal injury and the damage of centrifuge.
Don’t unpack and repair instrument by oneself. Please contact with our technicians for necessary matter. Attention for fireproofing Small breaker (air switch) can protect the circuit when centrifuge over-currents. Please use the breaker (air switch) with same specification and model when replacing it.
If centrifuge is irradiated by ultraviolet radiation for long time, the outer skin may be faded. Clothing should be covered for avoiding irradiation. This centrifuge must not be used to separate the sample which producing lammable and explosive gas. The stock around centrifuge is also forbidden. ...
Page 6
may be the disease-causing and poisonous matters. The centrifuge can not be manipulated with the matters unless you have toke the protection measures. Please pay attention to the liquid explanation in container before operation. Be careful of the infectious liquid. If not understanding the provided sample, please at least make sure that they aren’t microbe.
1. General introduction 1.1 Application and scope The centrifuge makes use of centrifugal theory to separate, purify and condense various mixture liquid or emulsion with tiny particle and similar density.It can be widely used in the fields of blood separation, pharmacy, biochemistry, genetic engineering and radioimmunogy ets.
2.2 Technical data 1. Max. Speed: 10000r/min 2. Max Capacity: 6×1000ml 3. Max. RCF(field): 13800xg 4. Time Range: 0~23h99min 5. Noise: ≤65dB 6. Temp Range: -20℃~+40℃ 7. Speed control accuracy±50 r/min 8. Ac/Dc Rte: 0-9 Grade 9. Program storage capacity: 35 Rotor data Rotor:According to the customer's request to process various rotor Number...
1). Display window: 2). Accelerate Key 3). Move up key 4). Confirm Key 5). Start Key 6). Decleration Key 7). Move down Key 8). Cancel Key 9). Stop/Open Key 3. Installation 3.1 Installation Requirements a). The centrifuge should be placed on the firm table with four feet contacting the table.
3.3 Rotor Installation First clasp the spoke of rotor with one hand, and put the solid wrench on the cap nut by another hand. In the mean while turn the wrench for 3 to 6 rounds in counterclockwise direction to loose the cap nut until the rotor can be separated from driving shaft.
3.5 Operation Procedures & Directions 3.5.1 Power on Power switch is located on the right bottom corner of control panel. Turn it to “On” position, the power indicator will turn red. 3.5.2 Parameters setting 1). After power on, the cooling system starts to work, now wait for 30 minutes for pre-cooling.
from 0-10000rpm), and step 9 is the lowest in speed up. When using the angle rotor is heavy, or using swing rotor, pls choose a slower grade, like 5-6. You are suggested to choose step 3 or 4. ⑥Dec setting: press ▲ or ▼button, the Dec display window is a digital flashing, adjust "+"...
when the speed display window does not show "0",pressing "open" key is invalid. 3.5.6 Note 1. The speed window shows the speed in running, but if you need to know the RCF, just perss ▲or ▼transfer to RCF , then the “Speed” display window will be a digital flashing, shows the RCF, press the "confirm"...
5. Trouble Shooting 1). The rotor does not work. a). If the power light is off, check if the fuse is burned out, whether the power contact is good. b). Whether the value set is correct. c). Whether it is under thermal protection after a long time working. 2).
remove the rotor. c). Do not run for a long time when there is no refrigeration. 6. Instrument safety and protection This centrifuge has multiple protection functions and will Stop automatically while trouble occurs. The error code will be displayed in the ‘SPEED/RCF’ window, then press [Stop] for display clearance.
While inverter has trouble, the system can not work and “E9” will be then displaying. Rotor disassembly:(after finishing adjust or centrifuge) First clasp the spoke of rotor with one hand, and put the solid wrench on the cap nut by another hand. In the mean while turn the wrench for 3 to 6 rounds in counterclockwise direction to loose the cap nut until the rotor can be separated from driving shaft.
Page 17
material defects or others caused by the factory. Accessories and technical service are to be provided at any time.
Quality Certificate Floortype High-Speed Refrigerated Centrifuge GL10-MC Max speed:10000r/min Max RCF:13800xg Serial No: It is certified that the centrifuge is passed quality inspection & test .
Warranty Card Name & Model Floortype High-Speed Refrigerated Centrifuge GL10-MC Specifications 6*1000ml Date of Delivery Address Service Records Date Contents Signature...
Service Manual Trouble Cause Removal No 220V power supply; Air switch to transformer is Check power supply turnoff(rosin joint) The display window not Power transformer in main board is damaged;Control Check and replace working board has burn out. The electrical elements L7812,L7805 of the display Replace board are broken.
Page 22
properly:light on means abnormal;light off means normal. The indicator light of inverter trouble:light on means LD5 small red light abnormal;light off means normal. The indicator light of inverter start:light on means LD6 small green light start;light off means stop. The indicator light of door lid opening : light on by LD7 small green light pressing door...
Page 23
Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co.,Ltd Add : Suite 1506,Lane581 XiuChuan Rd.,PuDong New Area,Shanghai,China Tel: 0086 21 50966080 Fax: 0086 21 50966080 Web : www.drawell.com.cn Email : sales01@drawell.com.cn...