IMMUNITY:
IEC 60601-1-2
Standard
Item
Test Levels for Home
Healthcare Environment
± 8 kV contact;
IEC 61000-4-
± 2 kV, ± 4 kV,
ESD
2:2008
± 8 kV, ± 15 kV air
10 V/m
IEC
80 MHz – 2.7 GHz
RS
61000-4-3:2010
80% AM at
1 kHz
± 2 kV
IEC
EFT
61000-4-4:2012
100 kHz
repetition frequency
IEC
± 0.5 kV,
Surge
± 1 kV
61000-4-5:2014
3V 0.15 MHz –
80 MHz
6V in ISM
IEC
CS
and amateur bands
61000-4-6:2013
between 0.15 MHz and
80 MHz
80% AM at 1 kHz
30 A/m
IEC
PFMF
50 Hz or
61000-4-8:2009
60 Hz
PowerDot
PD-01M
Remarks
Test Levels
± 2 kV,
± 4 kV, ± 8 kV
No
contact;
performance
± 2 kV,± 4 kV,
degradation
± 8 kV,
observed.
± 15 kV air
10 V/m
No
80 MHz –
performance
5.785 GHz
degradation
80% AM at
observed.
1 kHz
± 2 kV
No
performance
100 kHz
degradation
repetition
observed.
frequency
No
± 0.5 kV,
performance
± 1 kV
degradation
observed.
10V
No
0.15 MHz – 80
performance
MHz
degradation
80% AM at
observed.
1 kHz
No
± 0.5 kV,
performance
± 1 kV
degradation
observed.
58
Voltage Dips:
1) 0% UT; 0,5
cycle at 0º.
45º, 90º, 135º,
225º, 270º,
315º
2) 0% UT;
1 cycle;
Voltage
IEC
dips &
61000-4-
Single phase at 0º
voltage
11:2004
variations
3) 70% UT;
25/30 cycles;
Single phase at 0º
Voltage
Interruptions:
0% UT;
250/300
cycle;
Voltage Dips:
1) No performance
degradation
observed.
2) No performance
degradation
observed.
3) No performance
degradation
observed.
As on the
previous
Voltage
column
Interruptions:
Performance
degradation
(device stopped
charging) has been
observed only
during voltage
interruption
testing, but no
degradation
observed after
the testing.
59