PowerDot Uno Instruction Manual page 30

Table of Contents

Advertisement

IMMUNITY:
IEC 60601-1-2
Standard
Item
Test Levels for Home
Healthcare Environment
± 8 kV contact;
IEC 61000-4-
± 2 kV, ± 4 kV,
ESD
2:2008
± 8 kV, ± 15 kV air
10 V/m
IEC
80 MHz – 2.7 GHz
RS
61000-4-3:2010
80% AM at
1 kHz
± 2 kV
IEC
EFT
61000-4-4:2012
100 kHz
repetition frequency
IEC
± 0.5 kV,
Surge
± 1 kV
61000-4-5:2014
3V 0.15 MHz –
80 MHz
6V in ISM
IEC
CS
and amateur bands
61000-4-6:2013
between 0.15 MHz and
80 MHz
80% AM at 1 kHz
30 A/m
IEC
PFMF
50 Hz or
61000-4-8:2009
60 Hz
PowerDot
PD-01M
Remarks
Test Levels
± 2 kV,
± 4 kV, ± 8 kV
No
contact;
performance
± 2 kV,± 4 kV,
degradation
± 8 kV,
observed.
± 15 kV air
10 V/m
No
80 MHz –
performance
5.785 GHz
degradation
80% AM at
observed.
1 kHz
± 2 kV
No
performance
100 kHz
degradation
repetition
observed.
frequency
No
± 0.5 kV,
performance
± 1 kV
degradation
observed.
10V
No
0.15 MHz – 80
performance
MHz
degradation
80% AM at
observed.
1 kHz
No
± 0.5 kV,
performance
± 1 kV
degradation
observed.
58
Voltage Dips:
1) 0% UT; 0,5
cycle at 0º.
45º, 90º, 135º,
225º, 270º,
315º
2) 0% UT;
1 cycle;
Voltage
IEC
dips &
61000-4-
Single phase at 0º
voltage
11:2004
variations
3) 70% UT;
25/30 cycles;
Single phase at 0º
Voltage
Interruptions:
0% UT;
250/300
cycle;
Voltage Dips:
1) No performance
degradation
observed.
2) No performance
degradation
observed.
3) No performance
degradation
observed.
As on the
previous
Voltage
column
Interruptions:
Performance
degradation
(device stopped
charging) has been
observed only
during voltage
interruption
testing, but no
degradation
observed after
the testing.
59

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Uno and is the answer not in the manual?

This manual is also suitable for:

DuoPd-01m

Table of Contents