4.5 Inspection

Inspect the equipment for any damage such as crack, scratch or
deformation. Do not use it if any damage is found.

Packaging

4.6 Packaging
Pack the equipment in a sterile barrier such as Polypropylene fleece
or transparent package made from Polyethylene film and Tyvek®, and
then place it into a tray. The tray should be also covered with a sterile
barrier.
Additionally the equipment can be placed on plastic mesh wires supplied
for plasma sterilization and then packed as mentioned above.
The equipment can be packed in a simple or double packing.
Please note that the size of a sterile barrier should be large enough
to be able to pack the equipment leaving sufficient space to seal it
completely.
A sterile barrier should be sealed by an appropriate sealing machine
and it is important to confirm that the package is sealed completely.
If the sealing is not complete, pack and reseal again.
-10-
Q1E-EP0403