Ethernet/lxi remote high speed digitizer with 14/16 bit resolution (186 pages)
Summary of Contents for Spectrum 2172
Page 1
Setting the pH7 standard solution mode (sect. 3.4.3 of manual) The meter allows for calibration at a pH of either 6.8 (NIST), 7.0 (USA) or a customer- specified value. Spectrum ships the meter with pH 7.0 solution. To confirm the meter is in USA mode: 1.
Page 2
Calibration (sect. 2.5 of manual) 1. Open the internal solution filler port (fig. 3). 2. Immerse the tip of the probe into the pH7 to a depth of at least 3cm (fig. 4). Allow the read- ing to stabilize. 3. With the meter in pH measurement mode (this can be selected by pressing the MODE key), press the CAL key to enter calibration mode.
Page 3
Setting meter for Auto Temperature Compensation (sect. 3.4.4 of manual) 1. From Measurement mode, press the SET button until you’re in Temperature Compensation Setting Mode (fig. 9). 2. Press ENTER until the screen displays “Atc”. 3. Press MEAS to return to measurement mode. Figure 9 Set Cell Constant (sect.