Interacoustics Aqua Stim Instructions For Use Manual page 25

Water irrigator
Hide thumbs Also See for Aqua Stim:
Table of Contents

Advertisement

Guidance and manufacturer's declaration — electromagnetic immunity
The AquaStim is intended for use in the electromagnetic environment specified below. The customer or
the user of the AquaStim should assure that it is used in such an environment,
Immunity test
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the AquaStim is used exceeds the applicable RF compliance level above, the AquaStim should be
observed to verify normal operation, If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the AquaStim.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
D-0128227-A – 2020/12
Aqua Stim Water Irrigator – Instructions for Use - US
IEC / EN 60601 test
level
3 Vrms
150kHz to 80 MHz
3 V/m
80 MHz to 2,7 GHz
Compliance level
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
parts of the AquaStim, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
d
3 Vrms
d
3 V/m
d
GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
=
1
2 ,
P
1
2 ,
P
=
80 MHz to 800 MHz
2
3 ,
=
P
800 MHz to 2,7
b
a
Page 21