Page 1
UAM-820BT Arm Type Manufactured for Zewa, Inc. 12960 Commerce Lakes Drive#29 Thank you for selecting a ZEWA Blood Pressure Monitor UAM-820BT. Fort Myers, FL 33913 USA www.zewa.com Toll Free Customer Service: To use the monitor correctly and safely, please read the manual thoroughly.
CATALOGUE CATALOGUE Table of Contents INTRODUCTION........................2 General Description Indications for Use Contraindications Measurement Principle Safety Information LCD Display Signal Monitor Components List BEFORE YOU START......................9 The Choice of Power Supply Installing and Replacing the Batteries Pair-up the Blood Pressure Monitor with Your Device MEASUREMENT.........................12 Tie the Cuff Start the Measurement...
Then it starts inflating the cuff. Readings taken by the UAM-820BT are equivalent to those obtained by a Meanwhile, the unit detects pressure oscillation generated by beat-to-beat trained observer using the cuff and stethoscope auscultation method.
Page 4
INTRODUCTION INTRODUCTION CAUTION CAUTION * The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was * This device is intended for adult use in homes only. clinically investigated according to the requirements of ISO 81060-2:2013. * The device is not suitable for use on neonatal patients, pregnant women,patients with * To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, the manufacturer.
Systolic blood pressure * If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of ZEWA. Don’t open or repair the device by yourself in Low blood pressure Diastolic blood pressure the event of malfunctions.
Replace the batteries whenever the below happens 1. Blood Pressure Monitor 2. Cuff ( 23~37cm) 3. 4*AA Batteries shows (UAM-820BT) (Type BF applied part) The display dims The display does not light up CAUTION Do not use new and used batteries together.
CAUTION Interference may occur in the vicinity of equipment marked with the following symbol . And UAM-820BT may interfering vicinity electrical equipment. Sensitive people, including pregnant women pre-eclamptic and those who implanted medical electronic instruments, should avoid using the unit whenever possible.
Page 8
MEASUREMENT MEASUREMENT Start the Measurement Tie the Cuff 1.When the monitor is off,press the “ ” Remove all jewelry, such as watches and button to turn on the monitor, and it will finish bracelets from your left arm. the whole measurement. Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm.
MEASUREMENT DATA MANAGEMENT Recall the Records 2.This device will proceed to data transmission after measurement.The Bluetooth 1.When the monitor is off, please press and hold “ ” button to enter symbol blinks on the LCD indicates data is transmitting. the mode for Pair-up the Blood Pressure Monitor with Your Device. 2.
DATA MANAGEMENT DATA MANAGEMENT INFORMATION FOR USER Tips for Measurement CAUTION Measurements may be inaccurate if taken in the following The most recent record (1) is shown first. Each new measurement is circumstances. assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (60) is dropped from the list.
ABOUT BLOOD PRESSURE DATA MANAGEMENT What are systolic pressure and diastolic pressure? Maintenance When ventricles contract and pump blood out Systolic Diastolic In order to get the best performance, please follow the of the heart, the blood pressure reaches its blood entering blood discharging instructions below.
TROUBLESHOOTING INFORMATION FOR USER This section includes a list of error messages and frequently Why does my blood asked questions for problems you may encounter with your blood pressure monitor. If the product is not operating as you think it pressure fluctuate should, check here before arranging for servicing.
50hPa Measurement perimeter About 23cm~37cm of the upper arm For more information about our products or online support, please visit Weight www.zewa.com. To contact our customer service, please call 888-993-3592 or Approx.248g(Excluding the batteries) support@zewa.com. External dimensions Approx.102mm*143mm*73mm Attachment 4×AA batteries,user manual ,Carry bag...
COMPLIED STANDARDS LIST EMC GUIDANCE Complied Standards List EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Risk management Application of risk management to medical devices FCC Statement EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices. Symbols to be used with medical device labels, labelling and FCC ID:OU9TMB995-BS Labeling information to be supplied.
EMC GUIDANCE EMC GUIDANCE Table 2 EMC Guidance Guidance and manufacturer’s declaration – electromagnetic immunity 1)This product needs special precautions regarding EMC and needs to be The device is intended for use in the electromagnetic environment specified below. The customer or the installed and put into service according to the EMC information provided, user of the device should assure that it is used in such an environment.
Page 16
EMC GUIDANCE EMC GUIDANCE Table 3 Table 4 Guidance and manufacturer’s declaration – electromagnetic immunity Recommended separation distances between portable and mobile RF communications equipment and the device. The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Page 17
EMC GUIDANCE Table 5 Guidance and manufacturer’s declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device, should assure that it is used in such an environment. Radiated RF IMMUNITY Test...