Advertisement

Manufactured for Zewa, Inc.
12960 Commerce Lakes Drive#29
Fort Myers, FL 33913 USA
www.zewa.com
Toll Free Customer Service:
1-888-993-3592
warranty@zewa.com
User Manual
Blood Pressure Monitor
Thank you very much for selecting ZEWA Blood Pressure Monitor UAM-901
Please read the user manual carefully and thoroughly to ensure the safe usage of this
product. Keep the manual well for further reference in case you have problems.
Version:1.0
UAM-901
Arm Type

Advertisement

Table of Contents
loading

Summary of Contents for zewa UAM-901

  • Page 1 Manufactured for Zewa, Inc. 12960 Commerce Lakes Drive#29 Fort Myers, FL 33913 USA Thank you very much for selecting ZEWA Blood Pressure Monitor UAM-901 www.zewa.com Toll Free Customer Service: Please read the user manual carefully and thoroughly to ensure the safe usage of this 1-888-993-3592 product.
  • Page 2: Table Of Contents

    CATALOGUE CATALOGUE DATA MANAGEMENT......................31 Table of Contents Recall the Records Delete the Records INFORMATION FOR USER....................38 INTRODUCTION........................3 Tips for measurement General Description Maintenances Indications for Use Contraindications ABOUT BLOOD PRESSURE....................40 Measurement Principle What are systolic pressure and diastolic pressure? Safety Information What is the standard blood pressure classification? LCD Display Signal Irregular heartbeat detector...
  • Page 3: Introduction

    Contraindications Readings taken by the UAM-901 are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. 1.The device should not be used by any person who may be suspected of,or is pregnant .
  • Page 4: Safety Information

    INTRODUCTION INTRODUCTION CAUTION Safety Information * This device is intended for adult use in homes only. The signs below might be in the user manual, labeling or other component. * The device is not suitable for use on neonatal patients, pregnant women,patients with They are the requirement of standard and using.
  • Page 5 INTRODUCTION INTRODUCTION CAUTION CAUTION *Please check that operation of the device does not result in prolonged impairment of dose of a medicine prescribed by a doctor. Consult your doctor if you have any question patient blood circulation. about your blood pressure. * When measurement, please avoid compression or restriction of the connection tubing.
  • Page 6 INTRODUCTION INTRODUCTION CAUTION CAUTION the storage and operating conditions described in this booklet. from the arm and press the START/STOP button to stop inflation. * The equipment is not AP/APG equipment and not suitable for use in the presence of a * Before use, make sure the device functions safely and is in proper working condition.
  • Page 7 TURE. Otherwise, it may cause damage to the unit or danger to the user/patients. contact the SERVICE PERSONNEL of ZEWA. Don’t open or repair the device by * There is no luer lock connectors are used in the construction of tubing, there is a yourself in the event of malfunctions.
  • Page 8: Lcd Display Signal

    INTRODUCTION INTRODUCTION LCD Display Signal SYMBOL DESCRIPTION EXPLANATION Systolic blood pressure High blood pressure Diastolic blood pressure Low blood pressure Pulse display Pulse in beats per minute The average value of blood pressure Average value Indicate it is in the memory mode and Memory which group of memory it is.
  • Page 9: Monitor Components

    INTRODUCTION INTRODUCTION Monitor Components Component list of SYMBOL DESCRIPTION EXPLANATION pressure measuring system CUFF Blood pressure 1 Cuff Indicate the blood pressure level level indicator AIR HOSE 2 Air pipe Current Time Month/Day/Year, Hour/Minute 3 PCBA 4 Pump AIR CONNECTOR Blood pressure monitor is detecting a Heartbeat PLUG...
  • Page 10: List

    List The Choice of Power Supply 1.Blood Pressure Monitor 1 .Battery powered mode: 2.Cuff (Type BF applied part) (UAM-901) ( 22~42cm) 6VDC 4*AAA batteries AC adaptor powered mode: Please use the AC adaptor and USB cable just like the following picture(not included): 3.
  • Page 11: Installing And Replacing The Batteries

    BEFORE YOU START BEFORE YOU START CAUTION Installing and Replacing the Batteries Do not use new and used batteries together. Do not use different types of batteries together. • Slide off the battery cover. Do not dispose the batteries in fire. Batteries may explode or leak. •...
  • Page 12 BEFORE YOU START BEFORE YOU START 2 .Press “MEM” button to 4 .Repeat steps 2 and 3 to change the [YEAR]. set the [MONTH] and [DAY]. Each press will increase the numeral by one in a cycling manner. .Repeat steps 2 and 3 to 3 .When you get the right set the [HOUR] and [MINUTE].
  • Page 13: Select The User

    BEFORE YOU START BEFORE YOU START Then press “ MEM ” button again, select the user 6.After the [MINUTE] is set, ID between user 1 and user 2 . the LCD will display “donE” first,then display all the settings you have done and then it will turn off.
  • Page 14: Turn On Or Turn Off The Voice

    BEFORE YOU START BEFORE YOU START 2. Press “MEM ” button again to select turn on or turn off the voice. Turn on or turn off the voice It will speak “ Voice on” or “Voice off ” at the same time. 1.
  • Page 15: Measurement

    MEASUREMENT MEASUREMENT Tie the Cuff Helpful tips for Patients, especially for Patients with Hypertension: Remove all jewelry, such as watches and bracelets Rest for 5 minutes before first measurement. from your left arm.Note: If your doctor has diagnosed Wait at least 3 minutes between measurements. you with poor circulation in your left arm, use your right arm.
  • Page 16: Start The Measurement

    MEASUREMENT MEASUREMENT Start the Measurement Inflating and measuring. Display and save the results.(When the volume is on, 1 .When the monitor is off, press “Start/Stop” button to turn on START START the monitor, and it will finish the whole measurement . It will broadcast your measuring result.) STOP STOP...
  • Page 17: Data Management

    DATA MANAGEMENT DATA MANAGEMENT Recall the Records 1 . When the monitor is off, please press “MEM” button to show the recent record. 2 . Press “MEM” button If the records are less than or “SET” button to get 3 groups, it will display the the record you want.
  • Page 18: Delete The Records

    DATA MANAGEMENT DATA MANAGEMENT CAUTION 2.Press and hold “Start/stop” The most recent record (1) is shown first. Each new measurement is button to confirm deleting assigned to the first (1) record. All other records are pushed back one and the monitor will display “dEL donE+User ID”...
  • Page 19 DATA MANAGEMENT DATA MANAGEMENT B: Delete all records 2.Press and hold “Start/stop” button to confirm deleting 1.Hold pressing “MEM” button and the monitor will display for 3 seconds when the “dEL donE+USER ID” and then monitor is in the memory turn off.
  • Page 20: Information For User

    DATA MANAGEMENT INFORMATION FOR USER Tips for Measurement 3. If there is no record, Measurements may be inaccurate if taken in the following the right display will circumstances. show. Within 1 hour Immediate measurement Within 20 minutes after dinner or drinking after tea, coffee, smoking after taking a bath When talking or...
  • Page 21: About Blood Pressure

    INFORMATION FOR USER ABOUT BLOOD PRESSURE Maintenance What are systolic pressure and diastolic pressure? Systolic Diastolic When ventricles contract and pump blood out of the Please calibrate the blood pressure monitor in specific institute once heart, the blood pressure reaches its maximum value blood discharging blood entering every two years to ensure the precise measurement.
  • Page 22: Irregular Heartbeat Detector

    ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE Why does my blood CAUTION Please consult a physician if your measuring result falls outside the range. pressure fluctuate Please note that only a physician can tell whether your blood pressure value throughout the day? has reached a dangerous point.
  • Page 23: Troubleshooting

    ABOUT BLOOD PRESSURE TROUBLESHOOTING Why do I get a different What you need to pay This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the product is attention to when you measure blood pressure at home not operating as you think it should, check here before arranging for servicing.
  • Page 24: Specifications

    TROUBLESHOOTING SPECIFICATIONS Battery powered mode: The measurement E 03 shows Loosen the clothing on Error 6VDC 4*AAA batteries process does not the arm and then message Power supply AC adaptor powered mode: 5V 1A (Not included) detect the pulse measure again. (Please only use the recommended AC signal.
  • Page 25: Authorized Components

    SPECIFICATIONS AUTHORIZED COMPONENTS Authorized Components Storage & transportation Temperature:-4°F to +140°F (-20°C to +60°C) A relative humidity range of ≤ 93%, non-condensing, 1.Please use the manufacturer condition at a water vapour pressure up to 50hPa authorized AC adaptor (Adapter and USB cable are not included) Measurement perimeter About 8.66"~16.53"...
  • Page 26: Contact Information

    AUTHORIZED COMPONENTS COMPLIED STANDARDS LIST Complied Standards List Contact Information For more information about our products, please visit www.zewa.com.you can get EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Risk management Application of risk management to medical devices customer service, usual problems and customer download, Zewa will serve you anytime.
  • Page 27: Emc Guidance

    Warning:Portable RF communications equipment (including peripherals such as antenna cables and external usability engineering to medical devices antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment UAM-901, including Software life-cycle cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could...
  • Page 28 EMC GUIDANCE EMC GUIDANCE Table 2 Table 1 Guidance and manufacturer’s declaration – electromagnetic Immunity Guidance and manufacturer’s declaration - electromagnetic emissions Emissions test Compliance Immunity Test IEC 60601-1-2 Compliance level Test level RF emissions Group 1 Electrostatic ±8 kV contact ±8 kV contact CISPR 11 discharge (ESD)
  • Page 29 EMC GUIDANCE EMC GUIDANCE Table 3 0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°, 0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°, Voltage dips, short 180°, 225°, 270° and 315°.0 % UT; 180°, 225°, 270° and 315°. interruptions and Guidance and manufacturer’s declaration - electromagnetic Immunity 1 cycle and 70 % UT;...
  • Page 30: Warranty

    EMC GUIDANCE WARRANTY Radiated RF GSM 1800; Pulse 1720 1700- CDMA 1900; IEC61000-4-3 modulation b) 1990 GSM 1900; (Test 217Hz 1845 DECT; specifications LTE Band 1, 1970 ENCLOSURE 4,25; UMTS PORT IMMUNITY to Pulse Bluetooth, 2450 2400- RF wireless modulation WLAN, 2570 communica...
  • Page 31 WARRANTY WARRANTY...

Table of Contents