Emc Table - Advantech iPS-M S Series User Manual

Intelligent power system
Table of Contents

Advertisement

4.1 EMC Table

Guidance and manufacturer's declaration – electromagnetic emissions
The model IPS-M power system is intended for use in the electromagnetic environments
specified below. The customer or the user of the model IPS-M power system
should be assured that it is used in such an environment.
Emissions test
Conducted
Radiated
EN 60601-1-2
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations &
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The model IPS-M power system is intended for use in the electromagnetic environment
specified below. The customer or the user of the model IPS-M power system should assure
that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
iPS-M S series User Manual V.A1
Electromagnetic environment – guid-
Compliance
ance
The model IPS-M is suitable for use in all
establishments, including domestic estab-
lishments and those directly connected to
the public low-voltage power supply net-
work that supplies buildings used for
domestic purposes.
IEC 60601
Compliance level
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/
±1 kV for input/
output
output
lines
lines
±1 kV line(s) to
±1 kV line(s) to
line(s)
line(s)
±2 kV line(s) to
±2 kV line(s) to
earth
earth
3 A/m
3 A/m
20
Electromagnetic envi-
ronment – guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the rel-
ative humidity should be
at least 30 %.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typi-
cal commercial or hospi-
tal environment.