General Safety Instructions • Before use, the following section should be read carefully and in its entirety. • The N234-MI-1005 should only be installed and put into operation by qualified and trained personnel. • Do not open the housing. There are no serviceable parts inside.
The unit should not be operated in close proximity to atmospheres with combustible composites. • The N234-MI-1005 should also not be operated in close proximity to devices that emit electromagnetic radiation so as to avoid electromagnetic interference (e.g.
Page 4
When connecting the N234-MI-1005 Medical Isolator, note that: • You cannot connect any telecommunication appliances, voltage supplies or similar to the N234-MI-1005. By doing so, both the N234-MI-1005 and personnel could be harmed. • You can only connect network components complying with IEEE 802.3 or IEEE 802.3ab...
Intended Use The N234-MI-1005 is an appliance specially developed for use in medical technology. The purpose of the N234-MI-1005 is to connect in compliance with electrical safety standards IEC 60601-1 and EMC/IEC 60601-1-2 a medical appliance or system equipped with a network interface to a non-medical appliance or system (single PC or local network).
N234-MI-1005. Please instruct your network systems administrator or medicine technology department for this purpose. The installation of the N234-MI-1005 is performed by way of the two pluggable connections (RJ45 jacks). Plug the network connection cable into the socket of the N234-MI-1005 until the lock engages with an audible click sound.
5. Review the availability and sufficiency of the documents. Should any deficiency be found during a test, the N234-MI-1005 should not be used any longer, as the safety of the patients and the operator cannot be guaranteed. If in doubt, please contact your supplier or the manufacturer.
Page 8
Cleaning Use a dry or a moderately damp cloth to clean the N234-MI-1005. Make sure no fluids or moisture penetrates the device. 18-01-295-9337D2-EN.indd 8 2/6/2018 8:53:15 AM...
Specifications Type Designation MI 1005 Operating Mode Continuous Operation Operating Temperature 14 to 185°F / -10 to 85°C Operating Humidity 10% to 90% RH, Non-Condensing Operating Air Pressure 700 hPA to 1060 hPA Storage Temperature -13 to 185°F / -25 to 85°C Storage Humidity 10% to 95% RH, Non-Condensing Storage Air Pressure...
Warranty and Product Registration 1-Year Limited Warranty Tripp Lite warrants its products to be free from defects in materials and workmanship for a period of one (1) year from the date of initial purchase. Tripp Lite’s obligation under this warranty is limited to repairing or replacing (at its sole option) any such defective products.
Page 11
Visit www.tripplite.com/warranty today to register your new Tripp Lite product. You’ll be automatically entered into a drawing for a chance to win a FREE Tripp Lite product!* * No purchase necessary. Void where prohibited. Some restrictions apply. See website for details.
Page 12
1111 W. 35th Street, Chicago, IL 60609 USA www.tripplite.com/support 18-01-295 93-37D2_RevA 18-01-295-9337D2-EN.indd 12 2/6/2018 8:53:15 AM...