EXCELL PRECISION CO., LTD
5-3 Analogue Current / Voltage Output Interface (OP-03)
Location
1
2
A. Terminal (4 way)
1 : 4 ~ 20mA current output, A(+)
2 : 0 ~ 10V voltage output, V(+)
3 : Current / voltage signal, Com. (-)
4 : Ground / 0V, FG
B. SPAN adjustment
Current / voltage Span adjustment to increase value turn clockwise, decrease value turn
anticlockwise.
C. ZERO adjustment
Current / voltage Zero adjustment to increase value turn clockwise, decrease value turn
anticlockwise.
Analogue output interface specification
Resolution
Current output : 4 ~ 20mA ( 0 ~ 550 load)
Voltage output : 0 ~ 10V
040112XX
A
1
2
3
4
3
4
5
6
7
: 16 bits
50
B
1. A (+)
3. Com. ( - )
SPAN
ZERO
+
-
+
-
8
9
10
11 12 13
C
2. V (+)
4. FG
ON
CAL
ZSME400000004