Troubleshooting; Specifications - HoMedics THERA P BPS-050 Manual

Manual inflate blood pressure monitor
Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have
when the
run down.
START/STOP
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on
incorrectly.
display or the blood
pressure value is
Did you talk or move
displayed excessively
during measurement?
low (or high).
Shaking of the arm
with the cuff on.
Er error symbol shown
Did you move during
on display - measure-
measurement?
ment result exceeds
Shaking of the arm
measurement range.
with the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Data error.
on display
CA appears on display Recalibration alert.
29
CORRECTION
Replace them with
two new AA alkaline
batteries.
Re-insert the batteries
in the correct
positions.
Wrap the cuff
properly so that it is
positioned correctly.
Measure again.
Keep arm steady
during measurement.
Measure again. Do not
move during
measurement.
Check cuff connection.
Measure again.
Turn the unit off, then
measure again.
Remove and reload
the batteries.
Refer to page 25 for
more information.

SPECIFICATIONS

P P o o w w e e r r S S o o u u r r c c e e : :
Two 1.5V DC (AA) batteries
M M e e a a s s u u r r e e m m e e n n t t
Oscillometric
M M e e t t h h o o d d : :
Pressure: 40~250 millimeters mercury (mmHg)
M M e e a a s s u u r r e e m m e e n n t t R R a a n n g g e e : :
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
A A c c c c u u r r a a c c y y : :
Pulse: ±5% of reading
P P r r e e s s s s u u r r e e S S e e n n s s o o r r : :
Semi-conductor
I I n n f f l l a a t t i i o o n n : :
Manual Inflation
D D e e f f l l a a t t i i o o n n : :
Automatic Pressure release valve
M M e e m m o o r r y y C C a a p p a a c c i i t t y y : :
60 memories
A A u u t t o o - - s s h h u u t t - - o o f f f f : :
1 minute after last button operation
O O p p e e r r a a t t i i o o n n
Temperature: 50°F~104°F (10°C ~40°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 40~85% RH max.
S S t t o o r r a a g g e e
Temperature: 14°F~140°F (-10°C~60°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 10~90% RH max.
W W e e i i g g h h t t : :
.30 lbs/338 grams (without batteries)
Standard size adult arm cuff:
A A r r m m C C i i r r c c u u m m f f e e r r e e n n c c e e : :
9.5"–14.2" (24 – 36 cm)
4.50"(L) x 4.17"(W) x 2.25"(H)
D D i i m m e e n n s s i i o o n n s s : :
112mm(L) x 106mm(W) x 58mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
Note: These specifications are subject to change without notice.