[Error] EXCEEDED TORQUE LIMIT                        
[Warn] EXCEEDED JOINT LIMIT                           
[Warn] EXCEEDED ASSIST LIMIT 
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 
Causes
When this message is displayed and the 
 
device is stopped, the physical therapist 
may try to start the device again by 
pressing the "START" button on the tablet 
PC.  
If the problem persists, reboot the device 
and the tablet PC. If there is still a 
problem with the device, turn off the 
device immediately and contact the 
manufacturer. 
When the joint angle reaches these joint 
 
limits, the device ceases applying assistive 
force. Thereafter, the physical therapist 
may reset the device by asking the 
patient to stand with upright posture and 
tapping the "Set Stand Pose" button on 
the tablet PC. Pressing the "START" 
button will then cause the device to start 
providing assistive force again. 
Recommended action(s) 