8.2. Regular maintenance (Weekly Checklist)
Check that the bolts on the actuator cover are not loose.
Check that the bolts on the battery cover are not loose.
Check that the bolts on the thigh support frame are not loose.
8.3. Cleaning
Risk of electrical shock hazard due to moisture!
Electrical shock may be delivered to the patient and the user due to excessive moisture when the device is
turned on while cleaning.
‐
Do not clean while in use.
‐
Be sure to turn off the power before cleaning.
Property damage may occur due to improper cleaning procedures or agents!
The product may be damaged if it is cleaned and disinfected with improper procedures or agents.
‐
Use only this procedure for cleaning.
If the device is used by more than one patient, it is essential to clean the device with disinfectant before use.
To prevent cross‐infection between patients, the device (i.e., waist part, thigh support frames, and thigh support
straps) should be cleaned after each use. The device should be disinfected as follows:
① Dampen (but do not soak) a lint‐free cloth with purified water.
② Wipe the surface of the device for at least five (5) minutes.
③ Wipe the surface of the device with an (isopropyl) alcohol‐based disinfectant for at least five (5) minutes.
‐ Recommended Disinfectant: Cavi‐Wipes (Metrex Research, Kerr) or equivalent
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11)
Page 62 of 79