  • Page 1 Operating manual Original operating manual AVERO Motion Avero Motion & Avero Motion E Version 2.6/E Subject to technical modifications 2018-12-19...
  • Page 2: Table Of Contents

    The height adjustment................17 7.1.2 Reclining adjustment up / down ...............17 7.1.3 The hand shower (only AVERO Motion) ..........18 7.1.4 Turn on the handshower (only AVERO Motion E) ........18 7.1.5 Filling and filling height ................18 7.1.6 The display screens (only AVERO Motion) ..........19...
  • Page 3 BEKA Hospitec GmbH Power / Function LED Display (only AVERO Motion) ........20 Optional functions ...................20 Operation ......................20 Prior to use .....................20 During use ......................21 8.2.1 Using the AVERO Motion / AVERO Motion E ..........21 Cleaning/Disinfection ....................26 Cleaning the bath tub ..................26 Disinfecting the bath tub .................26...
  • Page 4: Imprint

    1.1 Acknowledgement Dear customer, we would like to express our sincere thanks for the trust you have placed in us by purchasing this BEKA Hospitec GmbH product. Our products are manufactured and tested according to stringent quality criteria. 1.2 Manufacturer’s address BEKA Hospitec GmbH Am Rübenmorgen 3...
  • Page 5: Introduction

    The maintenance, inspection, assembly and installation as well as well as further technical interventions on the product must only be executed by BEKA Hospitec either by specialised companies authorised to this effect by BEKA Hospitec. The operation of the product as well as technical interventions on the product must only be carried out by specially trained personnel.
  • Page 6: The Operating Manual

    3.3 Description In this manual, the product is called bath tub and (bath) tub. Operating manual AVERO Motion V2.6 E Page 6 of 48...
  • Page 7: Descriptions Of The Various Models

    AVERO Motion: 11 button control panel, tub spout and thermostat mixer Optional with 10 button hand control AVERO Motion E Without control panel, with 10 button hand control and attached fitting. Operating manual AVERO Motion V2.6 E Page 7 of 48...
  • Page 8: Safety

    The manufacturer BEKA Hospitec shall not accept any liability for any damages, failures or faults caused by improper operation or handling.
  • Page 9: List Of Used Safety Instructions

    CE-label in accordance with the EC-Directive on Medical Devices. Solely intended for indoor use. Protection class I Do not put feet on the frame. Protection against the penetration of liquids: Protection against IPX4 water splashing from all directions. Operating manual AVERO Motion V2.6 E Page 9 of 48...
  • Page 10: Warnings

    (current and potential compensation) must be checked for cracks in the cable. ▪ The grid connection must be done by a qualified electrician. The currently valid VDE regulations must be observed. Operating manual AVERO Motion V2.6 E Page 10 of 48...
  • Page 11: Transport

    To remove the packaging materials, you will need a cutter knife. NOTE Take care so as to not damage the product when using tools. Do not cut with the cutter in the cardboard. Operating manual AVERO Motion V2.6 E Page 11 of 48...
  • Page 12: Removing The Cardboard

    To connect the bath to the water supply, dirt filters must be used on site. Please contact your plumber to this effect. The bath tub must be connected through ½‘‘ hoses. Operating manual AVERO Motion V2.6 E Page 12 of 48...
  • Page 13: Electrical Connection

    The connection of the bath tub must be executed in accordance with the technical drawings issued by BEKA Hospitec either the guidelines for the set-up and layout of wet areas and the current VDE standards. All supply and drain hoses as well as the electrical cables must be installed so as to avoid the danger of tripping.
  • Page 14: First Start-Up

    Only when these conditions are met, the manufacturer considers himself responsible for the impact on the safety, the reliability and the function of the device. In the event of a new connection of a AVERO Motion bath tub, the technical data must be observed. WARNING To prevent scalding, the hot/cold water mixer must be set up in accordance with the local water supply prior to the first use.
  • Page 15: Operating Elements And Their Function

    Ball joint wall bracket Foot stand Handheld shower head Machine base + Handheld shower hose Door handle Disinfectant compartment Hand grip (1x left, 1x right) Color light system (optional) Air spa (optional) Operating manual AVERO Motion V2.6 E Page 15 of 48...
  • Page 16: Button Control Panel Avero Motion

    Operating elements and their function 7.1 11 button control panel AVERO Motion Description Description Height adjustment button UP Disinfection LED (yellow) Height adjustment button DOWN Disinfection button Air spa system button on/off Power/ Function LED Increase intensity of air spa...
  • Page 17: The Height Adjustment

    Operating elements and their function The operation of the seat and recliner tub AVERO Motion and AVERO Motion E is the same for both versions. If equipped with only the 10-key hand-held unit, the temperature displays, the disinfection button and the "hot and cold" warnings are dispensed with.
  • Page 18: The Hand Shower (Only Avero Motion)

    There are two fixed fill levels for filling the bath that can be selected. If the AVERO Motion is in the entry position (tilted) only the tub’s foot space will be filled. Press this button.
  • Page 19: The Display Screens (Only Avero Motion)

    Operating elements and their function Only for AVERO Motion: The bath filling is automatically switched off when the filling temperature is greater than 42 °C (scald protection, display flashes with "HOT"). If the scald protection activates (temperature display flashes "HOT"), the bath filling cannot be turned on using the button (13).
  • Page 20: Power / Function Led Display (Only Avero Motion)

    / or aroma and air spa system, please also refer to the enclosed operating manuals. 8 Operation AVERO Motion’s concept is to facilitate bathing or showering, to promote the patient’s mobility and provide the opportunity to enjoy a relaxing bath.
  • Page 21: During Use

    Fill the foot space of the AVERO Motion by pressing the button If your tub is equipped with the optional “automatic water stop” function, the filling will automatically stop when it reaches the fill level.
  • Page 22 The picture shows the extended, thus preventing the position of the locking door in the end position; AVERO Motion door from lever. the door can now be properly closing. If the door handle is not locked with the handle.
  • Page 23 Check water temperature with your Hand (without gloves) or with a thermometer! When the bath is finished, turn the rotary knob eccentric cam (overflow) in order to drain the water. Operating manual AVERO Motion V2.6 E Page 23 of 48...
  • Page 24 Note! Open the door only when there is no water in the tub. Open the AVERO Motion door. To open the door, unlock the door lock by turning the lever in a downward motion. Direction of lever...
  • Page 25 Please make use of the integrated disinfection (option), with which your AVERO Motion Movement is additionally equipped. The AVERO Motion bath tub is free standing on three sides and requires only little space. Operating manual AVERO Motion V2.6 E...
  • Page 26: Cleaning/Disinfection

    9.1 Cleaning the bath tub Clean the AVERO Motion with a soft, lint-free cloth, moistened with soapy water or a standard plastic cleaner. Please do not use abrasives to clean the bath tub! To avoid damages, no aerosol cleaners, sprays, abrasive cleaners or solvents must be used to clean the keypad.
  • Page 27: Checks/Tests

    Checks/tests 10 Checks/tests In order to ensure a safe use of our AVERO Motion bath tub and the protection of the users and the patients, BEKA Hospitec prescribes an annual safety check. The execution of the safety checks and maintenance must be documented and proven on request.
  • Page 28: Waste Disposal

    BEKA dealer, who will recycle the product in accordance with the locally applicable regulations and laws. For an environmentally-sound disposal, the company BEKA Hospitec GmbH will provide more information in its capacity of manufacturer. Please clean and disinfect the product prior to its disposal as well.
  • Page 29: Troubleshooting Help/After-Sales Service

    Check the seal for soiling and damages. a) Check water plug and if necessary readjust the screw of the water plug. The bathing water does not flow away. b) Check the seal for soiling and damages. Operating manual AVERO Motion V2.6 E Page 29 of 48...
  • Page 30 12 please contact the after-sales services of your dealer either the manufacturer. BEKA Hospitec GmbH Am Rübenmorgen 3 35582 Wetzlar Phone: +49(0)641-9 22 22-0 Fax: +49(0)641-9 22-22-20 info@beka-hospitec.de Operating manual AVERO Motion V2.6 E Page 30 of 48...
  • Page 31: Appendix

    10% to 80%, non- condensing 500 – 1100 hPa - Atmospheric Pressure: Batteries for Emergency Drainage: Battery type: Sealed lead acid battery Battery voltage 24 Volt DC Battery capacity: 1.2 Ah Operating manual AVERO Motion V2.6 E Page 31 of 48...
  • Page 32: Dimensions Avero Motion Tilt And Reclining Bath Tubs

    Appendix 13.2 Dimensions AVERO Motion Tilt and Reclining Bath Tubs Operating manual AVERO Motion V2.6 E Page 32 of 48...
  • Page 33: Wiring Diagram

    Appendix 13.3 Wiring Diagram 13.3.1 Connection at the side Operating manual AVERO Motion V2.6 E Page 33 of 48...
  • Page 34: Connection At The Rear

    Appendix 13.3.2 Connection at the rear Operating manual AVERO Motion V2.6 E Page 34 of 48...
  • Page 35: Connection At The Floor

    Appendix 13.3.3 Connection at the floor Operating manual AVERO Motion V2.6 E Page 35 of 48...
  • Page 36: Declaration Of Conformity

    Appendix 13.4 Declaration of Conformity Operating manual AVERO Motion V2.6 E Page 36 of 48...
  • Page 37: Accessories

    910150565 Comfort neck cushions 910150565 Comfort neck cushions 910150555 Manual operation 13.6 Spare parts / Consumables Spare parts and consumables are available upon request from your BEKA Hospitec distributor or directly from the manufacturer. Please note: You cannot change all parts yourself.
  • Page 38: Assembly Instructions

    The picture shows shower hose from the shower hose the elbow fitting on unscrewed from the the tub. elbow fitting Assembly of the shower hose is done in the reverse order! Operating manual AVERO Motion V2.6 E Page 38 of 48...
  • Page 39: Changing The Handheld Shower Holder

    Unscrew the unit unscrewed screw from the disinfectant hose disinfectant hose. unscrewed from the disinfectant shower head. Assembly of the handheld disinfectant shower head is done in the reverse order! Operating manual AVERO Motion V2.6 E Page 39 of 48...
  • Page 40: Changing The Disinfectant Hose

    The picture shows When fitting the drip plug into its casing, the removed ensure it is correctly locked. eccentric plugs Assembly of the drip plugs is done in the reverse order. Operating manual AVERO Motion V2.6 E Page 40 of 48...
  • Page 41: Electromagnetic Compatibility

    Guidance and manufacturer’s declaration - electromagnetic immunity (Table 201) The AVERO Motion been designed for use in the hereafter listed ELECTROMAGNETIC ENVIRONMENTS. The customer or the user of the AVERO Motion must ensure that the appliance is used in such environment.
  • Page 42 Guidance and manufacturer’s declaration - electromagnetic immunity (Table 202) The AVERO Motion been designed for use in the hereafter listed ELECTROMAGNETIC ENVIRONMENTS. The customer or the user of the AVERO Motion must ensure that the appliance is used in such environment.
  • Page 43 Guidance and manufacturer’s declaration - electromagnetic immunity (Table 204) The AVERO Motion been designed for use in the hereafter listed electromagnetic environments. The customer or the user of the AVERO Motion must ensure that the appliance is used in such environment. Immunity...
  • Page 44 Recommended distance between portable and mobile communications equipment and the AVERO Motion (Table 206) The AVERO Motion is intended for use in an electromagnetic environment with controlled HF interferences. The customer or the user of the AVERO Motion can avoid electromagnetic...
  • Page 45: Journal

    According the Medical Device Directive, you are compelled to keep a journal for this device. You can use this journal as template. Journal Device: AVERO Motion / AVERO Motion E Manufacturer: BEKA Hospitec GmbH, Am Rübenmorgen 3, 35582 Wetzlar Serial number: _____________________________________________ Date of purchase: _____________________________________________ Site: _____________________________________________...
  • Page 46 Appendix Periodic inspection, repair, DGUV-3, safety check, etc. Type of Date Result Measure Signature: inspection Operating manual AVERO Motion V2.6 E Page 46 of 48...
  • Page 47 Appendix Your notes …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… …………………………………………………………………………………………………… Operating manual AVERO Motion V2.6 E Page 47 of 48...
  • Page 48 BEKA Hospitec GmbH ●Am Rübenmorgen 3 ●35582 Wetzlar Germany● Phone/Fax:+49(0)6 41-9 22 20-0...