CALIBRATION
Update the TW's gain and offset calibration values with the following GPIB strings:
CAL{1,2,3}:MEAS:LOC:HIR:WATT:GAIN GAIN
CAL{1,2,3}:MEAS:LOC:HIR:WATT:OFFSET GAIN

3.27 REMOTE HIGH RANGE WATTS CALIBRATION

Connect the 48 ohm resistor to the output of the TW through the power analyzer. Set
up the power analyzer to measure watts.
Set up the TW to AC coupling, high range, local sense, 6.5 ampere current limit, output
relay closed and 108 volts with the following GPIB programming strings:
OUTP:COUP AC
SOUR:VOLT:RANGE HIGH
SOUR:SENSE REMOTE
SOUR{1,2,3}:CURR 6.5
OUTPUT ON
SOUR{1,2,3}:VOLT 108
Wait until the power analyzer has stabilized and record the watts reading as RDNG1.
Query the TW's watts using the GPIB string below, and record this value as MEAS1.
MEAS{1,2,3}:POW?
Program the TW to 268 volts AC using the string below:
SOUR{1,2,3}:VOLT 268
Wait until the power analyzer has stabilized and record the watts reading as RDNG2.
Query the TW's watts using the GPIB string below, and record this value as MEAS2.
MEAS{1,2,3}:POW?
Turn the TW's output off via the GPIB:
OUTP OFF
Calculate the gain and offset values for updating the TW:
GAIN = (RDGN2 – RDGN1)/(MEAS2 – MEAS1)
OFFSET = RDGN2 – GAIN * MEAS2
Update the TW's gain and offset calibration values with the following GPIB strings:
CAL{1,2,3}:MEAS:REM:HIR:WATT:GAIN GAIN
CAL{1,2,3}:MEAS:REM:HIR:WATT:OFFSET GAIN
3-38
TW SERIES

Advertisement

Table of Contents
loading

This manual is also suitable for:

Truewave tw3500Truewave tw1750

Table of Contents