CALIBRATION
Wait for the DMM to settle and record the value as RDGN1.
Send the following query command to the TW and record the return string as MEAS1:
MEAS{1,2,3}:VOLT?
Calculate the offset value:
VALUE1 = RDGN1 – MEAS1
Send the offset value to the TW:
CAL{1,2,3}:MEAS:REM:HIR:VOLT:OFFSET VALUE1
Measure the current offset by sending the following GPIB query and record the return
string as MEAS1:
MEAS{1,2,3}:CURR?
Calculate the offset value:
VALUE1 = -1.0 * MEAS1
Send the calculated offset to the TW:
CAL{1,2,3}:MEAS:REM:HIR:CURR:OFFSET VALUE1
Set the TW to 440 volts DC by sending the following GPIB string:
SOUR{1,2,3}:VOLT:OFFSET 440
Wait for the DMM to settle. Take a DMM reading and record the value as RDGN1.
Query the output voltage of the TW and record the return string as MEAS1:
MEAS{1,2,3}:VOLT?
Set the TW to -440 volts DC by sending the following GPIB string:
SOUR{1,2,3}:VOLT:OFFSET –440
Wait for the DMM to settle. Take a DMM reading and record the value as RDGN2.
Query the output voltage of the TW and record the return string as MEAS2:
MEAS{1,2,3}:VOLT?
3-10
TW SERIES