Summary of Contents for KLINGER US201

  • Page 1 Ultrasonic Heatmeter Operation Manual Ultrasonic Heatmeter Type US201 Operation Manual KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 2 Ultrasonic Heatmeter Operation Manual KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 3 Ultrasonic Heatmeter Operation Manual KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 4: Product Overview

    Clamp on type ultrasonic flow meter is easy to install and no need to cut off the pipe that saves you lots of troubles and cost. At the same time US201 has our unique calculate software to ensure the high accuracy and low velocity response. US201 ultrasonic flow meter widely application in oil industry, water treatment, pure water, chemical and etc.
  • Page 5: Specifications

    Encapsulated design. Transducer Standard / Maximum cable length: 30 ft / 1000 ft ( 9m / 305 m ). Weight 16*23*28cm, 3.2kg KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 6 Upstream sensor - Blue Downstream sensor Grounding Black Downstream sensor + Brown Downstream sensor - Blue I OUT+ 40~20mA Output I OUT- KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 7 The flow meter can always complete all tasks at the same time. The tasks (Including measurement, output, etc) will be carried out as usual, no matter in which display window. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 8: Keypad Functions

    To "Unlock" it, select "Unlock" in Window M54 and enter the original password. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 9: Quick Start

    0-80 deg. C. Press the “ Enter” to confirm. Note: Room temperature is 25 deg.C 。 WTMP 20 ( KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 10: Measurement Site Selection

    Choose a section of pipe that is always full of liquid, such as a vertical pipe with flow in the upward direction or a full horizontal pipe. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 11 Consider the inside condition of the pipe carefully. If possible, select a section of pipe where the inside is free of excessive corrosion or scaling. Choose a section of sound conducting pipe. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 12 Ultrasonic Heatmeter Operation Manual KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 13: Transducer Installation

    Side view Section Top View KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 14: Transducer Mounting Inspection

    4.2. Transducer Mounting Inspection Check to see if the transducer is installed properly and if there is an accurate and strong enough ultrasonic signal KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 15 (3) If the transducer is mounted at the pipe’s centerline on the same diameter, (4) If the scale is too thick or the pipe mounting is distorted in shape, etc. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 16 After the installation is complete, power on the instrument and check the result accordingly. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 17 The scale factor entered must be one that results from actual flow calibration. The scale factor can be input in Window M26. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 18: System Lock

    In Window M33-Range -UpperL (upper limit frequency output flow value), input 5000; In Window M33-Mode-Frange( frequency range), input 100、1000; In Window M33-Mode-Option, select “a.Flow Rate”; Typical OCT Output wiring diagram as below: KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 19: Totalizer Pulse Output

    Use “↑” and “↓” to switch. Calibrate the current loop 4mA output. Use an ammeter to measure the output current of current loop and adjust the displayed numbers at the same time. Watch the ammeter until it reads 4.00. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 20 Processing the data directly from the SD card file location on the PC could result in losing or destroying data if the SD card is removed while still being processed. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 21 The ESN, instrument types and versions are able to view in Window M50. ATTENTION Other Operating Refer to "6.2 Window Display Explanations". KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 22: Windows Display Explanations

    M27 Correction M28 SQA M30 Serial Port Parameter M31 AI Settings Input and Output Settings M32 CL Settings M33 OCT Settings KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 23 M61 Running Timer M62 CL Adjust Others M63 RTD Adjust M64 AI Adjust NOTE:The other menu features are retained by manufacturers. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 24: Display Explanations

    Display the Velocity. * Flow Rate and the Flow Total switch every 6 seconds, Use the ENTER to stop the switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 25 1000(E+3) x 10000(E+4) * Cool Rate and the Cool Total switch every 6 seconds, Use the ENTER to stop the switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 26 If it is not, it is necessary to check if the transducers are installed properly or if the parameters have been entered correctly. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 27 Once item 9 is selected, the relevant pipe sound velocity must be entered. to switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 28 Select the temperature of water. Temperature should be 0-80 deg. C. Press “Enter” to confirm. Note: Room temperature is 25 deg.C KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 29 4. Plus-In X Select transducer Mounting Methods Three mounting methods are available: 0. V Reflect method 1. Z Direct method 2. N KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 30 0.03, system will take all the measured flow velocity values from - 0.03 to + 0.03 as "0". Generally, 0.03 recommended most applications. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 31 Value Deviation =10 m Flow meter Display =250 m Normally, set the value as "0". Use “↑” and “↓” to switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 32 Select the energy totalizer value you want Reset Heat Cool Temperature Select Heat Input Options: 0. RTD 1. AI Use “↑” and “↓” to switch KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 33 The K factor is used to modify the measurement results. The user can enter a numerical value (other than "1") according to the actual calibration results. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 34 Engineer menu, suggest customer use the factory setting. Transducers power control Engineer menu, suggest customer use the factory setting. 0. Auto 1. Low 2. High KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 35 . 19200 None . 38400 None . 56000 None You can setting the order as following : 1-0: 3-2 0-1:2-3 3-2:1-0 2-3:0-1 KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 36 Set the CL output value according to the flow value at 4mA or 0MA. Set the CL output value according to the flow value at 20mA. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 37 Uart CTRL Select the OCT Range value. OCT check opinions a. Check 500 b. Check 1000 c. Check 3000 d. Check 5000 KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 38 Enter the Upper alarm value, any of the measured flow, which is higher than the high value, will activate the alarm in the OCT hardware or relay output signal. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 39 (US bbl) 7. Imperial Barrels (Imp bbl) 8. Oil Barrels (Oil bbl) The following time units are available: /Day /Hour /Min /Sec KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 40 10(E+1) f. x 100(E+2) h.x 10000(E+4) g.x 1000(E+3) Temperature Unit a. ℃ b. ℉ Use “↑” and “↓” to switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 41 Key Tone Use this menu to “ON”/ “OFF” the key tone. Language Setting the language of the flow meter 0.English 1.Chinese KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 42 Running Time With this function, it is possible to view the total Working days since the flow meter left the factory. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 43 The entire Menu which related to the temperature, cooling, heating, energy. Only display when it’s an energy meter. Flow meter doesn’t have the function. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 44: Error Diagnoses

    * Wait until liners solidified and saturated. Adjusting Gain * Adjusting gain for normal measurement. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 45: Frequently Asked Questions And Answers

    Check to see if "Set Zero" was carried out with fluid flowing inside the pipe (Refer to Window M22). If it is confirmed, recover the factory default in Window M22-Reset. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 46: Appendix 3 - Serial Interface Network Use

    RXD receive PIN 6 empty GND ground PIN 7 empty PIN 8 empty PIN 1 empty PIN9empty PIN 2 RXD send KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 47: Direct Connection Via Rs232 To The Host Device

    ±d.ddddddE±dd(cr) Note1 Return instantaneous velocity RVV(cr)(lf) ±d.ddddddE±dd(cr) Return positive accumulative flow RT+(cr)(lf) ±ddddddd.dE±d(cr) Note 2 RT-(cr)(lf) Return negative accumulative flow ±ddddddd.d±d(cr) KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 48 8-bit binary data is taken. E.g. The return information of the RT(cr)(lf) is :+1234567E+0m3(cr)(lf), (the relative binary system data is 2BH,31H,32H,33H,34H,35H,36H,37H,45H,2BH,30H,6DH,33H,20H,0DH, KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 49 2xN* N*x2 ( Data ) CRC ( Verify ) N* = Data register number 3. MODBUS Protocol function code 0x06 usage KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 50 Flow meter Address Function Code Register Address Register Number CRC Verify Code 4. Error Check The flow meter only returns one error code 0x02 which means data first address in error. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 51 Negative total – low word 32 bits int. $000C 40013 Negative total – high word $000D 40014 Negative total – exponent 16 bits int. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 52 Flow rate unit –char 1,2 String Note 1 $003E 40063 Flow rate unit –char 3,4 $003F 40064 Flow total unit – char 1,2 String KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 53 19200, 4 = 38400,5 = 56000 Notes: The following flow rate units are available: 0. "m3" -Cubic Meter -Liters 1. "l" -Gallons 2. "ga" KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 54 6. "Tn" 7. "kw" -Kwh 16 bits int—short integer, 32 bits int – long integer, 32 bits real—floating point number, String—alphabetic string KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 55: Appendix 6 - Flow Application Data

    2460 Glass 5970 Bronze 2270 Plastic 2280 Fiber glass-epoxy 3430 Polyethylene 1600 Glass 3276 PTFE 1450 Polyethylene 1950 Rubber 1600 2540 KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 56: Sound Velocity In Water (1 Atm) At Different Temperatures

    1511.3 1553.4 1543.9 1513.5 1553.7 1515.7 1554.0 Refer to the sound velocity of other fluids and materials, please contact the factory. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 57 SD card (datalogger) Sample: Ultrasonic Flowmeter (Energy version) w. Pulse, Relay, RS485, 4-20mA output, 2xPt1000 sensors, max 80grC, 9 m cable Code: US201-2-CD01-030-PT1000 KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
  • Page 58 KLINGER Danmark A/S Nyager 12-14 DK-2605 Broendby Denmark Phone +45 4364 6611 www.klinger.dk...