Clamp on type ultrasonic flow meter is easy to install and no need to cut off the pipe that saves you lots of troubles and cost. At the same time US201 has our unique calculate software to ensure the high accuracy and low velocity response. US201 ultrasonic flow meter widely application in oil industry, water treatment, pure water, chemical and etc.
Encapsulated design. Transducer Standard / Maximum cable length: 30 ft / 1000 ft ( 9m / 305 m ). Weight 16*23*28cm, 3.2kg KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 6
Upstream sensor - Blue Downstream sensor Grounding Black Downstream sensor + Brown Downstream sensor - Blue I OUT+ 40~20mA Output I OUT- KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 7
The flow meter can always complete all tasks at the same time. The tasks (Including measurement, output, etc) will be carried out as usual, no matter in which display window. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Choose a section of pipe that is always full of liquid, such as a vertical pipe with flow in the upward direction or a full horizontal pipe. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 11
Consider the inside condition of the pipe carefully. If possible, select a section of pipe where the inside is free of excessive corrosion or scaling. Choose a section of sound conducting pipe. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
4.2. Transducer Mounting Inspection Check to see if the transducer is installed properly and if there is an accurate and strong enough ultrasonic signal KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 15
(3) If the transducer is mounted at the pipe’s centerline on the same diameter, (4) If the scale is too thick or the pipe mounting is distorted in shape, etc. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 16
After the installation is complete, power on the instrument and check the result accordingly. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 17
The scale factor entered must be one that results from actual flow calibration. The scale factor can be input in Window M26. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Use “↑” and “↓” to switch. Calibrate the current loop 4mA output. Use an ammeter to measure the output current of current loop and adjust the displayed numbers at the same time. Watch the ammeter until it reads 4.00. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 20
Processing the data directly from the SD card file location on the PC could result in losing or destroying data if the SD card is removed while still being processed. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 21
The ESN, instrument types and versions are able to view in Window M50. ATTENTION Other Operating Refer to "6.2 Window Display Explanations". KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Display the Velocity. * Flow Rate and the Flow Total switch every 6 seconds, Use the ENTER to stop the switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 25
1000(E+3) x 10000(E+4) * Cool Rate and the Cool Total switch every 6 seconds, Use the ENTER to stop the switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 26
If it is not, it is necessary to check if the transducers are installed properly or if the parameters have been entered correctly. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 27
Once item 9 is selected, the relevant pipe sound velocity must be entered. to switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 28
Select the temperature of water. Temperature should be 0-80 deg. C. Press “Enter” to confirm. Note: Room temperature is 25 deg.C KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 29
4. Plus-In X Select transducer Mounting Methods Three mounting methods are available: 0. V Reflect method 1. Z Direct method 2. N KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 30
0.03, system will take all the measured flow velocity values from - 0.03 to + 0.03 as "0". Generally, 0.03 recommended most applications. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 31
Value Deviation =10 m Flow meter Display =250 m Normally, set the value as "0". Use “↑” and “↓” to switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 32
Select the energy totalizer value you want Reset Heat Cool Temperature Select Heat Input Options: 0. RTD 1. AI Use “↑” and “↓” to switch KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 33
The K factor is used to modify the measurement results. The user can enter a numerical value (other than "1") according to the actual calibration results. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 34
Engineer menu, suggest customer use the factory setting. Transducers power control Engineer menu, suggest customer use the factory setting. 0. Auto 1. Low 2. High KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 35
. 19200 None . 38400 None . 56000 None You can setting the order as following : 1-0: 3-2 0-1:2-3 3-2:1-0 2-3:0-1 KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 36
Set the CL output value according to the flow value at 4mA or 0MA. Set the CL output value according to the flow value at 20mA. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 37
Uart CTRL Select the OCT Range value. OCT check opinions a. Check 500 b. Check 1000 c. Check 3000 d. Check 5000 KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 38
Enter the Upper alarm value, any of the measured flow, which is higher than the high value, will activate the alarm in the OCT hardware or relay output signal. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 39
(US bbl) 7. Imperial Barrels (Imp bbl) 8. Oil Barrels (Oil bbl) The following time units are available: /Day /Hour /Min /Sec KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 40
10(E+1) f. x 100(E+2) h.x 10000(E+4) g.x 1000(E+3) Temperature Unit a. ℃ b. ℉ Use “↑” and “↓” to switch. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 41
Key Tone Use this menu to “ON”/ “OFF” the key tone. Language Setting the language of the flow meter 0.English 1.Chinese KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 42
Running Time With this function, it is possible to view the total Working days since the flow meter left the factory. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 43
The entire Menu which related to the temperature, cooling, heating, energy. Only display when it’s an energy meter. Flow meter doesn’t have the function. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Check to see if "Set Zero" was carried out with fluid flowing inside the pipe (Refer to Window M22). If it is confirmed, recover the factory default in Window M22-Reset. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 48
8-bit binary data is taken. E.g. The return information of the RT(cr)(lf) is :+1234567E+0m3(cr)(lf), (the relative binary system data is 2BH,31H,32H,33H,34H,35H,36H,37H,45H,2BH,30H,6DH,33H,20H,0DH, KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 50
Flow meter Address Function Code Register Address Register Number CRC Verify Code 4. Error Check The flow meter only returns one error code 0x02 which means data first address in error. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...
Page 51
Negative total – low word 32 bits int. $000C 40013 Negative total – high word $000D 40014 Negative total – exponent 16 bits int. KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk...