Troubleshooting; Power Supply; Usb-Related Operation - Nikon DS-Ri2 Instructions Manual

Microscope camera; monochrome microscope camera
Table of Contents

Advertisement

6

Troubleshooting

6
6
Troubleshooting
This chapter describes items to check if the DS-Ri2/DS-Qi2 does not function as you expected and measures to be taken.
Incorrect use might prevent the product from providing primary performance even though it is not damaged. If a symptom
such as those in the following table occurs, take action as described in the table before requesting repair.
If a problem that is not listed in the table occurs or if a problem that occurred is not resolved by taking action as described in
the table, disconnect the power cable of the device, and then contact your nearest Nikon representative.

Power Supply

6.1
6.1
Symptom
The power does not come on.
The device is hot when
touched.

USB-Related Operation

6.2
6.2
Symptom
Cannot be connected normally.
Cannot transmit data normally.
Possible cause
The AC adapter is not connected.
The power cord is connected
incorrectly to the AC adapter.
The power switch of the product is
turned off.
The system is used in a hot or confined
space.
Possible cause
The performance of the PC does not
satisfy the required specifications for
using this product.
A computer or OS which this product
does not support is used.
The USB device driver is not installed
correctly to the PC.
USB 3.0-certified products are not
used for each USB board, USB cable,
and USB hub.
Two USB hubs are used.
14
_____________________________________________________
Connect the DC output cable of the AC adapter for
the product to the DC IN 12V connector on the
product.
Use the specified power cord to plug the AC inlet of
the AC adapter into the AC power socket.
Flip the power switch to the "I" side to turn on the
power to the product.
Use the product within the operating environment
described in "Chapter 7 Major Specifications."
Note: When having an unusual odor such as a
burning smell, turn off the power switch
disconnect the power cord of the AC adapter,
and request repairs.
Use a PC satisfying requirements given in "Section
1.3 Peripheral Devices."
Use a PC with the following OS installed:
Windows 7
Note: Macintosh is not supported.
Install a USB host driver from Microsoft Corporation.
Note: Operation may be abnormal with the device
driver attached with the USB board.
To use the NIS-Elements, install the dedicated device
driver.
Use USB 3.0-certified products for USB board (on
PC), USB cable, and USB hub.
Keep a USB hub to one tier.
Action
Action

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds-qi2

Table of Contents