Product(s) : ALM500, ALM700, AKR550, AKR750, APH550, CS550, CSPOLA600
Implementation date : 2016.02.16
Criticality : normal
Please find below the connection procedures between ALM500/ALM700 (Focimeter) and AKR550/AKR750
(autokerato refractometer) with APH550 to import fitting data from the ALM and/or the AKR.
TECHNICAL INFORMATION
INSTRUMENTS
CONNECTION PROCEDURES
Essilor Instruments
64bis, avenue Aubert - 94306 Vincennes Cedex - France
Tél. : +33(0)1 72 94 71 00 - Fax : +33(0)1 72 94 70 27
International Customer Service
10
10
12
12
13
14
14
15
16
1/17
2
2
2
3
3
4
5
6
7
7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ALM500 and is the answer not in the manual?

Questions and answers

Summary of Contents for Essilor ALM500

  • Page 1: Table Of Contents

    2. Connect the APH550 system with a CSPOLA600 or a CS550 or a CP550 3. Connect the APH550 system with a AKR550 or a AKR750 4. Connect the APH550 system with a ALM500 or a ALM700 III. Settings 1. Configure the KB550 keyboard 2.
  • Page 2: Hardware

    RS232 DB9 male/female cable, for connecting ALM or AKR, ref. X83A56 or ref. X83A60 X83A60 Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 3: Connection Procedures

    2. plug in cable between the keyboard and the “Controller” port on the power supply unit > System APH550 is connected. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 4: Connect The Aph550 System With A Cspola600 Or A Cs550 Or A Cp550

    OR A Connection between the APH550 system and a CSPOLA or a CS550 or CP550 is infra-red. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 5: Connect The Aph550 System With Aakr550 Or Aakr750

    (ref. X83A60 or X83A56) between the AKR550 or the AKR750 and port “ref.” on the power supply unit Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 6: Connect The Aph550 System With Aalm500 Or Aalm700

    OR A To connect the system to the power supply unit: • plug in the cable (ref. X83A60 or X83A56) between the ALM500 or the ALM700 and the “Lensmeter” port on the power supply unit Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél.
  • Page 7: Settings

    When you hear the beep, remove your finger. The following screen will appears. > Go to the tab “Chart”. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27 7/17...
  • Page 8 Go to the tab “Communication”. Complete the fields “Baud rate LM” and “Baud rate ref.” with the same values as on the devices. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 9 REF corresponds to the AKR data importation • LM corresponds to the ALM data importation Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 10: Configure The Monitors And The Test Projector

    Select “APH550 Chanel” (channel) Select the same channel as selected on the APH550, A, B, C or D. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 11 Go to the menu “Chart Type” Select: • A: decimal • B: feet • C: metric Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27 11/17...
  • Page 12: Cp550

    • If the Output PC mode is on “B” then only communication with a PC is possible Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 13: Configuring An Alm500 Or An Alm700

    International Customer Service 4. C ALM500 ALM700 ONFIGURING AN OR AN Press the button to access the configuration menu. Press keys to access the “Data Output” menu, in order to configure the serial link settings. Press the key The following screen is displayed: >...
  • Page 14: Exporting Data From An Alm Or An Akr

    International Customer Service IV. E XPORTING DATA FROM AN OR AN 1. E ALM500 ALM700 XPORTING DATA FROM AN OR AN Center a lens on the ALM. Refer to the relevant note if needed. Take a measurement. You can take a measurement on a lens or a frame.
  • Page 15: Exporting Data From An Akr750 Or An Akr550

    During data export, the message “Data transfer” is displayed on the screen. When the transfer has finished, the message “Data output success” is displayed on the screen. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 16: Importing Data On The Aph550

    The menu “Import data from” is displayed: > This menu allows you to import data manually or automatically. Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27...
  • Page 17 > Press “SET” to set up the corresponding lenses in the head. Written by D. Biasot Checked by H Rizki Approved by F. Drouard Essilor Instruments 64bis, avenue Aubert – 94306 Vincennes Cedex – France Tél. : +33(0)1 72 94 71 00 – Fax : +33(0)1 72 94 70 27 17/17...

This manual is also suitable for: