Quick start – Rear panel
(31) Z INPUT
This input is used to modulate the trace brightness via an external signal.
A (+) signal reduces the brightness; a (-) signal increases it.
(32) YA OUTPUT
Replica of the YA 20 mV/div. input signal, e.g. for feeding into a frequency
meter.
(33) FUSE HOLDER / VOLTAGE SELECTOR (see 2.1. Installation)
(34) POWER SOCKET
3
Operation – Instructions sheet (see
attachment)
Instructions Sheet: Oscilloscope
Use the oscilloscope in display and voltmeter mode.
Shunt connection
Before: All pushbuttons in up position (pulled out)
SCREEN BRIGHTNESS and ILLUMINATION low
VERTICAL and HORIZ POSITION at mid range
TRIGGERING on AUTO and INT
VARIABLE knobs in CAL position
Oscilloscope power-up
Settings:
reasonable BRIGHTNESS
SHARPNESS on the finest trace
VERTICAL YA lever on O, adjust POSITION (trace on axis)
HORIZ POSITION as required
How to set the oscilloscope
1st question
What do we want to
display and
measure?
22
1 voltage
YA = f(t)
YA = f(t)
and YB = f(t)
Press on YA
Press on YA and YB
or
2 voltages
YB = f(YA)
XY
Set YB
Set X and Y
BRIGHTNES
S