W A R R A N T Y Each CytoTherm product is produced under rigid quality control standards. This unit is fully warranted for a period of two years from the date of purchase. Call 609 396-1456 or 800 747-9699 for help. If necessary send unit to: CytoTherm 110 Sewell Ave.
The CT-4T comes with one CORRAL that will thaw 1 to 6 plasma bags at a time. A second CORRAL will let you thaw 6 more plasma bags for a total of up to 12 plasma bags at a time. The operator never comes in contact with the water in the bath when handling the plasma bags.
Page 3
Make sure that the DRAIN Valve is closed (horizontal). Fill the BATH with tap water to the bottom of the SHELF if you have one CORRAL mounted and to the screw of the shelf if you have 2 CORRALs mounted. If your tap water contains a lot of lime, use distilled or de-ionized water, add a pinch (0.01gr.) of salt to make the water conductive so that the level sensors will work.
OPERATION Wait for the water to heat up to 37.0 For your protection, use gloves when working with any blood products. Flat-frozen plasma units can be thawed without an overwrap bag. Hold a bag of flat-frozen plasma (bare or overwrapped) with the ports up and the label on the plasma unit away from the center of the CORRAL.
ENTER, the heater will heat continuously and the pumps will turn on. When the temperature reaches 1.0C above the target temperature the alarm will sound. Turn Off the unit. If alarm does not sound, contact CytoTherm. Alarm test should be performed every 3 months as part of unit maintenance.
Page 6
TOO HOT is an audible and displayed alarm condition that indicates that the bath temperature is 1.0C higher than the set temperature. If the operator pours hot water into the bath he/she will trigger this alarm. This should never happen in normal operation. If it does, contact CytoTherm.
Page 7
Turn Off the unit. If alarm does not sound, contact alarm does not sound, contact CytoTherm. Alarm test should be performed every 3 months as part of unit maintenance. CytoTherm. Alarm test should be performed every 3 months as part of unit maintenance.
Page 8
PARTS LIST CT-4T 14-MP Pump, Magnetic DP-SW Switch, 2-pole 14-77 Heater (2 needed) C-TH Safety Thermostat CT4-PC Printed Circuit Board C4-DISP Display C4T-BTH Bath Section C4-UMB Wiring Harness C-RM Rocking Motor C4T-PM Program Module C4-TS Temperature Sensor C4T-CLM Clamp ( bag to cor.)
Page 9
SAFETY FEATURES WATER LEVEL SENSOR The unit will display LOW WATER and beep if there is not enough water to cover the level sensor inside the bath, a screw located 3/4" above the temperature sensor. The heat will also turn off. The water must be conductive in order for the level sensor to work.