Table 2
Guidance and manufacturer's declaration – electromagnetic Immunity
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
IEC 61000-4-8
Conduced RF
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601-1-2
Test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
50 Hz / 60 Hz
Not applicable
10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
50 Hz / 60 Hz
Not applicable
10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz