Device Specification

Power supply
Display mode
Measurement mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
External dimensions
Mode of operation
Degree of protection
Protection against
ingress of water
Device Classification
Software Version
Weight
WARNING: No modification of this equipment is allowed.
Battery powered mode: 6V DC 4x AA batteries
AC Adaptor powered mode: 6V
(Please only use the recommended AC Adaptor).
Digital LCD V.A. 3.1in*3.6in (78mm*92mm)
Oscillographic testing mode
Rated cuff pressure:
0 mmHg ~ 299 mmHg (0 kPa ~ 39.9 kPa)
Measurement pressure:
SYS: 60 mmHg ~ 230 mmHg (8.0 kPa ~ 30.7 kPa)
DIA: 40 mmHg ~ 130 mmHg (5.3 kPa ~ 17.3 kPa)
Pulse value: (40-199) beat/minute
Pressure:
41°F-104°F (5°C-40°C) within ± 3mmHg (0.4kPa)
Pulse value: ±5%
A temperature range of: +41°F to +104°F
(+5°C to +40°C) A relative humidity range of
15% to 90%, non-condensing, but not requiring a
water vaporpartial pressure greater than 50 hPa
An atmospheric pressure range of:
700 hPa to 1060 hPa
Temperature: -4°F-140°F (-20°C to +60°C)
A relative humidity range of ≤ 93%, non-condensing,
at a water vapor pressure up to 50hPa
About 8.6in-16.5in (22cm~42cm)
Approx. 393g (Excluding the batteries)
Approx. 6.1in*4.8in*2.7in (154.3mm*121.5mm*68.1mm)
Continuous operation
Type BF applied part
IP21 means the device is protected from touch and
objects greater than 12.5mm, and protect
against vertically falling water drops.
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
A01
1A (not included)