Abbott FreeStyle Libre 2 User Manual page 216

Flash glucose monitoring system
Hide thumbs Also See for FreeStyle Libre 2:
Table of Contents

Advertisement

Glucose Reading Availability
The System is designed to show a Sensor glucose reading after each scan that is performed
throughout the wear period after the start-up time. Tables 8a and 8b show the glucose reading
capture rate for each day of the wear duration.
Table 8a: Glucose Reading Capture Rate Over Wear Duration (Adult; n=146)
Day of Wear
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Number of Sensors
146
145
143
140
138
135
134
131
128
123
120
113
112
104
Capture Rate (%)
98.3
98.1
98.3
98.3
98.4
98.3
98.4
98.4
98.4
98.4
98.4
98.5
98.5
98.6
210