Abbott FreeStyle Libre 2 User Manual page 212

Flash glucose monitoring system
Hide thumbs Also See for FreeStyle Libre 2:
Table of Contents

Advertisement

Table 6a: Sensor Accuracy Relative to YSI over the wear duration (Adult; n=144)
Number of
Wear
CGM-reference
Period
Beginning
Early Middle
Late Middle
End
Table 6b: Sensor Accuracy Relative to YSI over the wear duration (Pediatric*; n=129)
Number of
Wear
CGM-reference
Period
Beginning
Early Middle
Late Middle
End
* Includes children 6-17 years of age. No YSI measurements were obtained for children 4-5 years of age.
MARD (%)
pairs
6955
9.9
4522
8.5
3503
8.8
3755
9.1
MARD (%)
pairs
1828
10.7
1642
8.0
1534
9.7
1542
10.2
Within ±15% /
Within ±20% /
±15 mg/dL
±20 mg/dL
83.4
87.7
86.8
86.4
Within ±15% /
Within ±20% /
±15 mg/dL
±20 mg/dL
79.6
89.5
83.6
82.6
Within ±40% /
±40 mg/dL
90.4
99.3
94.5
99.8
93.4
99.7
92.9
100.0
Within ±40% /
±40 mg/dL
88.5
98.6
94.2
98.5
92.9
99.5
91.1
99.3
206