Table 2
Guidance and manufacturer's declaration – electromagnetic emission
Immunity test
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV, ±8 kV, ±15 kV, air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV signal input/output
IEC 61000-4-4
100 kHz repetition frequency
Surge
±0.5 kV, ±1 kV differential mode
IEC 61000-4-5
±0.5 kV, ±1 kV,±2 kV common mode
Voltage dips, short
0%UT; 0.5 cycle. At 0°, 45°, 90°, 135°,
interruptions and
180°,225°,270° and 315°. 0%UT ; 1 cycle
voltage variations on
and 70%UT ; 25/30 cycles. Single phase: at
power supply input
0°. 0% UT ; 300 cycle
lines
IEC 61000-4-11
Power frequency
30 A/m
magnetic field
50Hz/60Hz
IEC 61000-4-8
Conducted RF
3 V
IEC61000-4-6
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
10 V/m
IEC61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
17 I GB
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV, air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
0%UT; 0.5 cycle. At 0°, 45°, 90°, 135°,
180°,225°,270° and 315°. 0%UT ; 1 cycle
and 70%UT ; 25/30 cycles. Single phase: at
0°. 0% UT ; 300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz