RANGE? QUERY
Returns the selected range.
Syntax:
Response:
0 = RANGE_OFF
1 = R20mOHM
2 = R200mOHM
3 = R2_OHM
4 = R20_OHM
5 = R200_OHM
6 = R2K_OHM
7 = R20K_OHM
A = RANGE_AUTO
RDNG? QUERY
Responds with reading from the device in engineering notation. Query function also uses
an immediate update of the conversion so repeated query requests gets the most
accurate data.
Syntax:
Example:
Response:
Power-on default = 0.000e+0
CAUTION: Will not work if the transmit interrupt happens to be busy with a display update
and the following occurs:
Remote
System
System
System
System
Remote
System
Remote
System
System
System
RANGE?
Range number = 0 – 7
(SAFE MODE)
RDNG?
RDNG?<crlf>
2.4321e+1<crlf>
<value in reduced engineering notation<crlf>>
->RANGE 6
<2kΩ)
->Places "REM^" in TX buffer (Tiny Display)
->Change to R6
->Places "2k^" in TX buffer (Tiny Display)
->Formats Display String "1.6543k^"
->READING?
->Places in TX buffer to VFD 'D' characters
->RANGE 4
->Change to R4
->Executes READING? query, uses Range=4 and last displayed value
->Places in TX buffer, next 'R' characters
50
(on 20.000Ω ranges)