GE Healthcare Life Sciences Instructions 52-1777-00 AM Filter Kit HR10/HR16 Filter Kit HR 10, Filter Kit HR 16 These instructions describe how to change filters and O-rings on HR 10 and HR 16 columns.
Page 2
Table of Contents Filter kit contents ..................When to change the filter ..............How to change the filter ................ Troubleshooting ..................52-1777-00 AM...
Page 3
1 Filter kit contents These filter kits (Code Nos. 18-3575-01/18-3585-01 respectively) contain 10 filters and 4 O-rings. 2 When to change the filter Change the filter on the top of your column when you observe an increased backpressure, a space which has become visible between adapter and filter, a loss of resolution and always before you perform a thorough column cleaning.
Page 4
Step Action Carefully and slowly bring up the loose filter by using the hook (part of the Filter tool). Do not press or puncture the medi- um bed. Note: Only a very small amount of medium should be in suspension after removing the filter.
Page 5
Step Action Check the O-ring in the adapter. Change it if it is damaged or asymmetrical. Do not use a sharp tool to remove the O-ring, as this could damage the adapter and cause leakage. Insert the adapter, aligning it so that it engages the slots in the end piece.
Page 6
Use a flow rate of 1 to 1.5 ml/min for 10 to 30 minutes. For further information contact your local GE Healthcare represen- tative. 52-1777-00 AM...
Page 7
All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information.