Siemens POWERMOBIL Installation And Setting Instructions Manual page 5

Hide thumbs Also See for POWERMOBIL:

Advertisement

Area dose product measurement device
Potentiometer
Fig. 1
Calibrating the area dose product measuring device
NOTICE
CONSTANT parameters = T
Read and record the
measuring device.
Programming
Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
Start the service program.
Select the "Adjustment, Parameter..." menu.
Select "Diamentor" in the "Parameter Groups" combo box.
Click on "Get from Unit".
The parameter "1. Diamentor Existing" must be on YES.
Select parameter "2. Constant
Enter the parameter
Click on "Set Value".
Click on "Put to Unit".
Click on "Save to File".
NOTICE
Siemens AG
Medical Engineering
This adjustment is used to calibrate the area dose product mea-
suring device to the POWERMOBIL/ ARCOSKOP display.
w
T
value shown on the electronics unit of the area dose product
w
T
.
w
T
in the "Value Actual:" field.
w
You must exit the service program with "Logoff" and "Quit" in
order to save the modified parameters to diskette as a new file.
SPR2-240.814.04
Rev. 02
Page 5 of 10 POWERMOBIL/ARCOSKOP
07.00
TD SD 24