Area dose product measurement device
Potentiometer
Fig. 1
Calibrating the area dose product measuring device
NOTICE
CONSTANT parameters = T
•
Read and record the
measuring device.
Programming
•
Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
•
Start the service program.
•
Select the "Adjustment, Parameter..." menu.
•
Select "Diamentor" in the "Parameter Groups" combo box.
•
Click on "Get from Unit".
•
The parameter "1. Diamentor Existing" must be on YES.
•
Select parameter "2. Constant
•
Enter the parameter
•
Click on "Set Value".
•
Click on "Put to Unit".
•
Click on "Save to File".
NOTICE
Siemens AG
Medical Engineering
This adjustment is used to calibrate the area dose product mea-
suring device to the POWERMOBIL/ ARCOSKOP display.
w
T
value shown on the electronics unit of the area dose product
w
T
.
w
T
in the "Value Actual:" field.
w
You must exit the service program with "Logoff" and "Quit" in
order to save the modified parameters to diskette as a new file.
SPR2-240.814.04
Rev. 02
Page 5 of 10 POWERMOBIL/ARCOSKOP
07.00
TD SD 24